Maintenance and Servicing
16
112037
6
Maintenance and Servicing
6.1
General Requirements
Repairs of the
biovac 106 Aspiration Systems
may only be performed by the manufacturer
or authorized workshops. The prerequisites are a complete and factually correct damage re-
port, and a clean and, if necessary, a decontaminate device.
Send in defective devices for repair either to the manufacturer or to an authorized workshop.
The information about the contamination or thorough cleaning is legally binding parts of the
contract.
6.2
Maintenance Performed by the User
WARNING !
Only perform the work that is described here, and that which is permitted to be done
by the user.
All other maintenance and service work may only be performed by the manufacturer
or a dealer authorized by him.
Beware of the pump parts being possibly contaminated by hazardous substances.
Wear protective clothing if there is contamination.
The operator may perform maintenance work to the extent indicated below:
WARNING !
Before opening the pump unplug it from the mains.
6.2.1
Maintenance of the aspiration systems
6.2.1.1
Disassembly
Disconnect the power supply by pulling the device plug.
Disconnect the hose connections.
To empty, clean and decontaminate the secretion container, lift it up out of the holder of
the aspiration system.
Unscrew the suction connection and exhaust air filter.
Unscrew from below the four inner screws on the device.
Lift the hood off. Now, you can start with the maintenance of the diaphragm pump accord-
ing
to chapter 6.2.2
.
WARNING !
Wear protective gloves or protective clothing if necessary!
Do not clean with compressed air!