Description
8
112037
3
Description
3.1
Design
►Aspiration System biovac 106 with secretion glass container 2 liters
A filter holder for a PTFE filter is located between the secretion glass container
(1)
and the
diaphragm pump
(2)
. Filters down to a minimum pore size of 0.22 µm may be inserted in this
holder. According to the filter fineness used, it is possible to achieve different types of filtra-
tion, enabling, among other uses, the
handling of bacteria and viruses.
The cap
(3)
of the secretion glass container consists of a plastic cap with a seal.
Fig. 1a Aspiration-System biovac 106 with secretion glass container 2 liters
A quick coupling
(4)
is used for the connection to the diaphragm pump
(2)
. It is provided with
a stop valve.
The used-resistant diaphragm pump
(2)
used has a long useful life.
Secretion container
(1)
and handvac
(5)
are connected with a hose
(6)
. After disconnecting
the unions
(7)
, the two flexible tube ends at the container may be connected to one another
to ensure safe transport.
The Aspiration System is switched on and off directly at the device switch or the optionally
available foot switch
(9)
.
The user appropriate pipette tips
(8)
can be used in the respective adapter of the handvac
(5)
. All hoses
(6)
located outside of the Aspiration System are made of silicone.
(9)
Option
(2)
(4)
(3)
(6)
(5)
(8)
(7)
(1)
(6)
(6)