Installation and Operation
112037
13
5
Installation and Operation
5.1
Unpacking
Carefully unpack the
biovac 106 Aspiration System
.
Check for:
Transport damage,
Conformity with the specifications of the supply contract (type, electrical supply data),
Completeness of the delivery.
Please inform us without delay if there are discrepancies between the delivery and the con-
tractually agreed scope of delivery, or if damage is detected.
Please take note of the general terms of business of the manufacturing firm.
In case of a claim under warranty, the device must be returned in packaging that is
suitable for protecting it during transport.
5.2
Installation and Connection
Procedure:
Fig. 5 Rear side - connections
(biovac 106
with secretion glass container 2 liters)
1.
Set the aspiration system on a flat and
horizontal surface.
2.
Connect the “handvac” to the suction line
(A)
.
(see also in appendix operation manual)
3.
Connect the exhaust pipe to the
building ventilation system
(B)
.
4.
If required, connect the foot switch
(accessories optional)
to the socket
(C)
provided on the unit.
5.
Set the mode selector switch
(D)
to
"Hand" or "Foot".
(see chapter 3.5)
6.
Voltage selector 115/230 V
(E)
Label on device:
Instructions examine and label remove!
7.
Connect the system to the power
supply
(F)
.
8.
Switch on the unit's main switch
(G)
.