![Gamma Vacuum DIGITEL Instruction Manual Download Page 30](http://html.mh-extra.com/html/gamma-vacuum/digitel/digitel_instruction-manual_3370463030.webp)
Table 8. Gamma protocol commands
Hex
Command Description
Data
Field
Response
Data/Response
Description
1
SYS GET MODEL
Get controller model
string.
DIGITEL TSPq
2
SYS GET FIRMWARE
VERSION
Get firmware version.
SW Version
X.XX
Where:
X.X is firmware
version number
4F
SYS TFTP SERVER ADDRESS
Configures the IP address
of the TFTP server used in
the firmware upgrade
process.
If no parameter is
specified, the current
address is returned.
X.X.X.X
X.X.X.X
Where:
X.X.X.X is the IP
address of the TFTP
server
8F
SYS SET FIRMWARE
UPDATE
Tells system firmware
update is wanted. Send
the command and cycle
power. Upon power up,
bootloader shall be
running.
No parameters.
FF
SYS RESET
Restarts the system.
N
Where:
N is the reboot mode
0 = Reboot
1 = Reboot and start
bootloader app
58
HV GET/SET DIGITAL
INPUT
Configures digital inputs.
If only N parameter is
specified, the current
digital input configuration
is returned.
N, S, F
S, F
Where:
N is the digital input
(1-4)
S must be set to 0
F is function:
“00” - Off
“03” – TSP 1 Interlock
“04” - TSP 2 Interlock
28
TSP TURN OFF
Turns off TSP sublimation.
R
Where:
R is TSP (1-2)
29
TSP SET ACTIVE FILAMENT
Sets TSP active filament.
R, N
Where:
R is the TSP (1-2)
N is the filament
number (1-4)
2D
TSP TURN ON
R
Where:
Page 30