![Gamma Vacuum DIGITEL Instruction Manual Download Page 25](http://html.mh-extra.com/html/gamma-vacuum/digitel/digitel_instruction-manual_3370463025.webp)
Pin
Description
27
Not connected
28
Not connected
29
Reserved - do not use, ground
30
ANALOG OUT 1 - output, analog, range 0 to +10 V d.c.
31
Ground
32
ANALOG OUT 2 - output, analog, range 0 to +10 V d.c.
33
Ground
34
ANALOG OUT 3 - output, analog, range 0 to +10 V d.c.
35
Ground
36
ANALOG OUT 4 - output, analog, range 0 to +10 V d.c.
37
DIGITAL OUT 4 - output, digital, 0/+5 V d.c., 7 mA maximum
* Maximum relay current 500 mA, maximum voltage 28 V. Common and NC
pins are connected in default or de-energized state.
Table 2. J112 Connector pinout
RS-232
RS-485 Full Duplex (four
wire)
RS-485 Half Duplex (two
wire)
RXD
2
+TX
2
+TX/+RX
2
TXD
3
-TX
8
-TX/-RX
8
GND
5
GND
5
GND
5
+RX
3
-RX
7
Cabling for RS-232 Serial Protocol:
The controller functions as Data Terminal Equipment (DTE) devices. When
the controller is connected to another DTE device (such as a personal
computer), a null modem serial cable or a simple three wire serial crossover
cable is required to connect the devices. The null modem cable swaps the
signal lines so the receive and transmit signals are properly connected.
Note:
For-RS 232, the serial cable between the PC and the controller only needs to cross Rx
and Tx pins. No other signals are needed, except ground pin 5.
Page 25
900035_B
900035_B - Rear panel