Filament - Selects active filament (1-4). This filament is used in the
sublimation process.
The filament status is displayed in brackets next to filament number.
Filament status reads (ok) if at the end of the last sublimation cycle, filament
current reached at least 80% of the target current value.
The status will read (open?) if filament current did not reach 80% of the
target current value.
Current Target (A) - Configures sublimation current target level.
On Time (s) - Configures filament sublimation time in seconds.
Number of Cycles - Configures number of sublimation cycles to run.
Cycle period - Configures time in minutes between each sublimation cycle.
Cycle period is not applicable (N/A) when number of cycles is set to 1.
TSP1/TSP2 - Selects TSP for which above parameters are configured.
7.3.
Display setup
Figure 10. Display setup screen
Brightness Level - Configures screen brightness level (1-10).
Screensaver - Configures screen-saver on/off state. Turning screensaver off
is not recommended as this will shorten screen backlight lifetime. When
screen-saver is enabled/active, display brightness level is reduced to level 1.
Touchscreen calibration - Invokes touchscreen calibration process.
Page 19
900035_B
900035_B - Touchscreen user interface