Current target (A) - Sublimation current target level.
On time (s) - Cycle sublimation time in seconds.
Number of cycles - Number of sublimation cycles to run.
Cycle period - Time in minutes between each sublimation cycle. Not
applicable when number of cycles is 1.
Advance options configure - Invokes TSP advance options window.
Figure 17. TSP setup screen
9.16.
TSP pressure window setup
Available setup:
Pressure window - Turns on and off pressure window option.
When pressure window is enabled sublimation process will run only if pump
pressure is within the defined pressure window. If pump pressure is outside
the defined pressure window, the TSP status will read ‘TSP Armed, Outside
the pressure window’ and sublimation process is not allowed.
Pressure high limit - Specifies upper pressure limit.
Pressure low limit - Specifies lower pressure limit.
Supply source - Specifies which supply and pump pressure reading will be
taken into consideration for pressure window.
Page 30
900031_J
900031_J - Touchscreen user interface