Pump size (L/s) - Configures pump size (0 - 1200).
Auto restart - Configures supply auto HV restart at controller power up (on
and off).
Output voltage (V) - Configures output voltage target level (3 - 7 kV).
Current limit (mA) - Configures output current limit (2 - 125 mA).
Configure advanced options - Invokes advance options setup.
Figure 4. Supply setup screen
9.3.
Supply setup advanced options
Available setup:
Main display value - Configures home screen large value (current, voltage, or
pressure).
Pressure factor - Configures pump pressure factor.
Configure foldback (off) - Configures and enables or disables the foldback
function. When foldback function is enabled and pump pressure falls below
defined foldback pressure point, the output voltage is reduced to define
foldback output voltage.
Figure 5. Supply setup advanced options
Page 20
900031_J
900031_J - Touchscreen user interface