background image

Pressure:

Off

Low

Med

High

Off

Low

Med

High

Off

Low

Med

High

Off

Low

Med

High

When selecting a 

program: 

Sleep

Time

Time

Run

Pressure

Off

Low

Med

High

55/13

50/10

45/7

40/5

ºF/ºC

Target

Temperature

Actual 

Temperature

Displays pressure.

Displays amount of 
time unit will run.

Displays amount of 
time unit will sleep.

oPerAtiNG tHe sYsteM 

to operate your GrPro 2.1 system, you need: 

  Control Unit filled with ice and water as indicated by the fill line labels within the 

reservoir. Optimal performance is achieved by first adding 1.5 liters of water, and then 

filling the reservoir to the top with ice.

  Game Ready supplied power supply.  

  Connector Hose.  

  Wrap (includes a Heat Exchanger and Sleeve).  

Notes:

  The Wrap is comprised of an inner Heat Exchanger and an outer Sleeve. The combination 

of Sleeve and Heat Exchanger is referred to throughout this manual as a “Wrap.” To ensure 

proper performance, be sure to use ATX Series Heat Exchangers.

  The GRPro 2.1 Control Unit should be placed on a stable surface (such as the floor or a 

table) during use.   

 

  Note that using the System in an environment with a high ambient temperature may 

affect its ability to provide adequate cooling, or may limit the ice life.   

 

  If you will be using the System with a Game Ready optional battery pack (sold separately), 

please consult the instructions for use that accompany that battery pack. 

Warning: 

  Your GRPro 2.1 Control Unit should be plugged into a grounded electrical outlet prior to 

operation.

  Position the Control Unit to minimize the risk of tripping over the Control Unit, Connector 

Hose, or power cord.

Precautions: 

  Failure to properly follow the instructions of this manual and those of your medical 

provider may interfere with or prevent delivery of appropriate therapy. 

  To avoid risk of electrical shock unplug the Control Unit from the electrical outlet prior to 

filling the Control Unit with ice and water.

1

  Push the door release button to open the ice box door.

When the system 
is running in 
Program Mode:

          

Displays the amount of run time 
remaining.

Run time count-down bars.

Displays the amount of sleep time 
remaining.

When the system 
is running in 
Manual Mode:

          

Displays the amount of run time 
remaining.

Sleep time  
count-down bars.

EN-5

EN-4

SUPERSEDED

Summary of Contents for GRPro 2.1

Page 1: ...odel Numbers 550550 03 550550 03 RN 550550 53 GRPro 2 1 Control Unit User s Manual53 Game Ready Control Unit User s Manual PN 703758 Rev D 2011 CoolSystems Inc All rights reserved Brought to you by Co...

Page 2: ...of Operation______________________________________ 2 Buttons_________________________________________________ 2 Adjusting Temperature___________________________________ 3 Display______________________...

Page 3: ...n on Programs Play Pause Use this button to start or pause a treatment Add Time Use this button to add time in Manual Mode does not work in Program Mode You can add up to 90 minutes Treatment must be...

Page 4: ...a table during use Note that using the System in an environment with a high ambient temperature may affect its ability to provide adequate cooling or may limit the ice life If you will be using the S...

Page 5: ...Make sure you hear it click To disconnect simply press the blue or gray button and remove the connector from the Wrap 9 Set your target temperature displayed in the Target Temperature window by turnin...

Page 6: ...id your warranty Abrasive materials to clean the Control Unit Doing so will damage the plastics and will void your warranty Caution The Control Unit is not a waterproof device Do not apply a direct st...

Page 7: ...tomer Service at 1 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor 1 Identify the filter within the ice reservoir 2 Using two fingers grasp and squeeze the tw...

Page 8: ...e in the acute stages of inflammatory phlebitis in the affected region Who have any history or risk factors for deep vein thrombosis or pulmonary embolus including prolonged bed rest in the affected r...

Page 9: ...armful ingress of water 5 3 This product provides ordinary protection against ingress of water Degree of safety in the presence of flammable anesthetics or oxygen 5 5 Not suitable for use in an oxygen...

Page 10: ...ufacturer s Declaration Electromagnetic Immunity The GRPro 2 1 is intended for use in the electromagnetic environment specified below The customer or the user of the GRPro 2 1 should assure that it is...

Page 11: ...g NO Pressure wrap shall be vented to atmosphere The cooling temperature of the circulating ice water will be adjustable between 34 F 1 C and 50 F 10 C as long as the ice water in the ice box is suppl...

Page 12: ...e connections Disconnect and reconnect Wrap from Connector Hose Make sure there is water in the ice box Verify that the ice box filter is not clogged Make sure there are no kinks in the Wrap or Connec...

Reviews: