background image

error

WHAt does  

it MeAN?

WHAt CAN  

i do?

under Pressure: 

The 

Control Unit can’t reach 

its target maximum 

compression. This often 

indicates that there is 

a leak in the pneumatic 

compression circuit, either 

in the Connector Hose, 

Wrap or Control Unit. Or 

it may occur because the 

hook and loop fastener on 

your Wrap has worn out.

• 

 Make sure the Wrap is applied securely.

• 

 Try using a different Wrap and Hose to 

isolate which component may be producing 

the error. For example, an error which 

occurs with a Shoulder Wrap but not a Knee 

Wrap may indicate that the Shoulder Wrap 

is causing the error: not the Control Unit. 

• 

 If using a Dual Hose, make sure you have 

two Wraps attached.

deflation error:

 The Control 

Unit has detected that 

the Wrap has not properly 

deflated.

• 

 Turn the Control Unit off.

• 

 If there is air left in the Wrap, disconnect 

the Wrap from the Connector Hose and 

manually deflate the Wrap by applying 

pressure to it.

• 

 Reconnect the Hose to the Wrap and 

reapply the Wrap to the body.

• 

 Turn the Control Unit on and press play/

pause.

• 

 Make sure the Wrap is applied securely 

against the body.

• 

 If possible, try using a different Wrap and 

Hose to isolate which component may be 

producing the error. For example, an error 

which occurs with a Shoulder Wrap but not a 

Knee Wrap may indicate that the Shoulder 

Wrap is causing the error: not the Knee 

Wrap or Control Unit.

Pump Performance error:

 

The Control Unit has 

determined that the fluid 

pump may be working too 

hard. This could be caused 

by ice or debris in the fluid 

circuit. In order to prevent 

possible damage to the fluid 

pump, the unit will stop 

therapy.

• 

 Turn the Control Unit off and back on again.

• 

 Reapply the Wrap, making sure to follow 

all application instructions accompanying 

the Wrap.

• 

 Disconnect and reconnect the hose from 

the Control Unit and the Wrap, verifying 

that an audible “click” is heard at both 

connection points.

• 

 If that does not solve the problem turn the 

Control Unit off for 20 minutes (to let the 

pump cool down) before turning it on again 

to try again.

low flow:

 Control Unit 

has detected something 

blocking the water flow.

• 

 

Be sure you are using ATX Series Heat 
Exchangers.

• 

 Check all hose connections.

• 

 Disconnect and reconnect Wrap from 

Connector Hose.

• 

 Make sure there is water in the ice box.

• 

 Verify that the ice box filter is not clogged.

• 

 Make sure there are no kinks in the Wrap or 

Connector Hose.

• 

 Reapply the Wrap snugly, making sure 

to follow all application instructions 

accompanying the Wrap.

• 

 Turn the Control Unit off and on again.

• 

 Disconnect and reconnect the hose from 

the Control Unit and the Wrap.

EN-20

PERIOD. NO WARRANTIES APPLY AFTER THAT PERIOD. SOME JURISDICTIONS DO 
NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, SO THIS 
LIMITATION MAY NOT APPLY TO YOU.

liMitAtioNs of liAbilitY

COOLSYSTEMS’ RESPONSIBILITY UNDER THIS, OR ANY OTHER WARRANTY, IMPLIED 
OR EXPRESS, IS LIMITED TO REPAIR OR REPLACEMENT, AS SET FORTH ABOVE. 
THESE REMEDIES ARE THE SOLE AND EXCLUSIVE REMEDIES FOR ANY BREACH 
OF WARRANTY. COOLSYSTEMS IS NOT RESPONSIBLE FOR DIRECT, SPECIAL, 
INCIDENTAL, OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY BREACH OF 
WARRANTY OR UNDER ANY OTHER LEGAL THEORY INCLUDING, BUT NOT LIMITED 
TO, LOST PROFITS, DOWNTIME, GOODWILL, AND DAMAGE TO OR REPLACEMENT 
OF EQUIPMENT AND PROPERTY. SOME JURISDICTIONS DO NOT ALLOW THE 
EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES, SO 
THE ABOVE LIMITATIONS OR EXCLUSIONS MAY NOT APPLY TO YOU. THIS LIMITED 
WARRANTY GIVES YOU SPECIFIC RIGHTS, AND YOU MAY ALSO HAVE OTHER RIGHTS 
THAT VARY FROM JURISDICTION TO JURISDICTION.

W A r r A N t Y   r e G i s t r A t i o N

reGistrAtioN 

Please complete the Warranty Registration for both the GRPro Control Unit and the 
Wraps online at www.gameready.com. You must register within 30 days from the date of 
purchase in order to receive warranty service.
 
You will need the following information to complete your Warranty Registration: 

the 

Control unit’s model number (ref) and its serial number (sN).

 These numbers are 

located on the label on the bottom of the Control Unit. Simply go to www.gameready.
com, click on the Product Registration link at the top of the page, fill out and submit your 
information.

exteNded WArrANties 

Extended Warranties are available for the GRPro System. For details and information, in 
the U.S. call Game Ready Customer Service at 1.888.426.3732 (+1.510.868.2100); from 
outside of the U.S. please contact your local distributor.

EC

REP

EMERGO EUROPE

Molenstraat 15

2513 BH, The Hague

The Netherlands

0086

EN-23

SUPERSEDED

Summary of Contents for GRPro 2.1

Page 1: ...odel Numbers 550550 03 550550 03 RN 550550 53 GRPro 2 1 Control Unit User s Manual53 Game Ready Control Unit User s Manual PN 703758 Rev D 2011 CoolSystems Inc All rights reserved Brought to you by Co...

Page 2: ...of Operation______________________________________ 2 Buttons_________________________________________________ 2 Adjusting Temperature___________________________________ 3 Display______________________...

Page 3: ...n on Programs Play Pause Use this button to start or pause a treatment Add Time Use this button to add time in Manual Mode does not work in Program Mode You can add up to 90 minutes Treatment must be...

Page 4: ...a table during use Note that using the System in an environment with a high ambient temperature may affect its ability to provide adequate cooling or may limit the ice life If you will be using the S...

Page 5: ...Make sure you hear it click To disconnect simply press the blue or gray button and remove the connector from the Wrap 9 Set your target temperature displayed in the Target Temperature window by turnin...

Page 6: ...id your warranty Abrasive materials to clean the Control Unit Doing so will damage the plastics and will void your warranty Caution The Control Unit is not a waterproof device Do not apply a direct st...

Page 7: ...tomer Service at 1 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor 1 Identify the filter within the ice reservoir 2 Using two fingers grasp and squeeze the tw...

Page 8: ...e in the acute stages of inflammatory phlebitis in the affected region Who have any history or risk factors for deep vein thrombosis or pulmonary embolus including prolonged bed rest in the affected r...

Page 9: ...armful ingress of water 5 3 This product provides ordinary protection against ingress of water Degree of safety in the presence of flammable anesthetics or oxygen 5 5 Not suitable for use in an oxygen...

Page 10: ...ufacturer s Declaration Electromagnetic Immunity The GRPro 2 1 is intended for use in the electromagnetic environment specified below The customer or the user of the GRPro 2 1 should assure that it is...

Page 11: ...g NO Pressure wrap shall be vented to atmosphere The cooling temperature of the circulating ice water will be adjustable between 34 F 1 C and 50 F 10 C as long as the ice water in the ice box is suppl...

Page 12: ...e connections Disconnect and reconnect Wrap from Connector Hose Make sure there is water in the ice box Verify that the ice box filter is not clogged Make sure there are no kinks in the Wrap or Connec...

Reviews: