table 2 for transient electromagnetic immunity

guidAnCe And mAnufACturer’s deCLArAtiOn –  

eLeCtrOmAgnetiC immunitY

The GRPro

 2.1 is intended for use in the electromagnetic environment specified below. The 

customer or the user of the GRPro

 2.1 should assure that it is used in such an environment.

immunitY tEst

iEC 60601 tEst lEvEl

ComplianCE lEvEl

ElECtromagnEtiC EnvironmEnt – guidanCE

Electrostatic

discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered with

synthetic material, the relative humidity

should be at least 30%.

Electrical fast

transient/burst

IEC 61000-4-4

± 2 kV for power 

supply lines
± 1 kV for input/

output lines

± 2 kV for power 

supply lines
± 1 kV for input/

output lines

Mains power quality should be that 

of a typical commercial or hospital 

environment.

Surge

IEC 61000-4-5

± 1 kV line(s) to 

line(s)
± 2 kV line(s) to 

earth

± 1 kV line(s) to 

line(s)
± 2 kV line(s) to 

earth

Mains power quality should be that 

of a typical commercial or hospital 

environment.

Voltage dips, short

interruptions and

voltage variations

on power supply

input lines

IEC 61000-4-11

<5% 

U

T

(>95% dip in 

U

T

)

for 0.5 cycle
40% 

U

T

(60% dip in 

U

T

)

for 5 cycles
70% 

U

T

(30% dip in 

U

T

)

for 25 cycles
<5% 

U

T

(>95% dip in 

U

T

)

for 5 s

<5% 

U

T

(>95% dip in 

U

T

)

for 0.5 cycle
40% 

U

T

(60% dip in 

U

T

)

for 5 cycles
70% 

U

T

(30% dip in 

U

T

)

for 25 cycles
<5% 

U

T

(>95% dip in 

U

T

)

for 5 s

Mains power quality should be that 

of a typical commercial or hospital 

environment. If the user of the GRPro

 

2.1 requires continued operation 

during power mains interruptions, it is 

recommended that the GRPro

 2.1 be 

powered from an uninterruptible power 

supply or a battery.

Power frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields 

should be at levels characteristic of a 

typical location in a typical commercial 

or hospital environment.

NOTE 

U

T

 is the AC mains voltage prior to application of the test level.

EN-16

table 3 for rf electromagnetic immunity

guidAnCe And mAnufACturer’s deCLArAtiOn – eLeCtrOmAgnetiC 

immunitY

The GRPro

 2.1 is intended for use in the electromagnetic environment specified below. The 

customer or the user of the GRPro

 2.1 should assure that it is used in such an environment.

immunitY tEst

iEC 60601 tEst lEvEl

ComplianCE lEvEl

ElECtromagnEtiC EnvironmEnt – guidanCE

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3Vrms

3 V/m

Portable and mobile RF communications 

equipment should be used no closer to any 

part of the GRPro

 2.1, including cables, 

than the recommended separation distance 

calculated from the equation applicable to 

the frequency of the transmitter.

recommended separation distance

d

 = 1.2 √

P

   80 MHz to 800 MHz

d

 = 1.2 √

P

   80 MHz to 800 MHz

d

 = 2.3 √

P

   800 MHz to 2.5 GHz

where 

P

 is the maximum output power 

rating of the transmitter in watts (W) 

according to the transmitter manufacturer 

and 

d

 is the recommended separation 

distance in meters (m).

Field strengths from fixed RF transmitters, 

as determined by an electromagnetic site 

survey,

a

 should be less than the compliance 

level in each frequency range.

b

Interference may occur in the vicinity of 

equipment marked with the following 

symbol:

   

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

a

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 

cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to 

fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field 

strength in the location in which the GRPro

 2.1 is used exceeds the applicable RF compliance level 

above, the GRPro

 2.1 should be observed to verify normal operation. If abnormal performance is 

observed, additional measures may be necessary, such as re-orienting or relocating the GRPro

 2.1.

b

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

EN-17

SUPERSEDED

Summary of Contents for GRPro 2.1

Page 1: ...odel Numbers 550550 03 550550 03 RN 550550 53 GRPro 2 1 Control Unit User s Manual53 Game Ready Control Unit User s Manual PN 703758 Rev D 2011 CoolSystems Inc All rights reserved Brought to you by Co...

Page 2: ...of Operation______________________________________ 2 Buttons_________________________________________________ 2 Adjusting Temperature___________________________________ 3 Display______________________...

Page 3: ...n on Programs Play Pause Use this button to start or pause a treatment Add Time Use this button to add time in Manual Mode does not work in Program Mode You can add up to 90 minutes Treatment must be...

Page 4: ...a table during use Note that using the System in an environment with a high ambient temperature may affect its ability to provide adequate cooling or may limit the ice life If you will be using the S...

Page 5: ...Make sure you hear it click To disconnect simply press the blue or gray button and remove the connector from the Wrap 9 Set your target temperature displayed in the Target Temperature window by turnin...

Page 6: ...id your warranty Abrasive materials to clean the Control Unit Doing so will damage the plastics and will void your warranty Caution The Control Unit is not a waterproof device Do not apply a direct st...

Page 7: ...tomer Service at 1 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor 1 Identify the filter within the ice reservoir 2 Using two fingers grasp and squeeze the tw...

Page 8: ...e in the acute stages of inflammatory phlebitis in the affected region Who have any history or risk factors for deep vein thrombosis or pulmonary embolus including prolonged bed rest in the affected r...

Page 9: ...armful ingress of water 5 3 This product provides ordinary protection against ingress of water Degree of safety in the presence of flammable anesthetics or oxygen 5 5 Not suitable for use in an oxygen...

Page 10: ...ufacturer s Declaration Electromagnetic Immunity The GRPro 2 1 is intended for use in the electromagnetic environment specified below The customer or the user of the GRPro 2 1 should assure that it is...

Page 11: ...g NO Pressure wrap shall be vented to atmosphere The cooling temperature of the circulating ice water will be adjustable between 34 F 1 C and 50 F 10 C as long as the ice water in the ice box is suppl...

Page 12: ...e connections Disconnect and reconnect Wrap from Connector Hose Make sure there is water in the ice box Verify that the ice box filter is not clogged Make sure there are no kinks in the Wrap or Connec...

Reviews: