Manual Mode:
Set the time in five minute increments by pushing the +/- buttons.
Set the pressure by pushing the pressure button. You can select
from 4 pressure settings: No pressure, Low Pressure (5-15mmHg),
(Medium Pressure (5-50mmHg), High Pressure (5-75mmHg).
Program Mode:
Push the program button to enter Program Mode. In Program Mode,
the unit will operate continuously according to the selected program.
You will need to replenish ice and water as previously indicated in
step 2.
You can select from the following programs: (Push the Program
Button to scroll through the available programs.)
Program 1:
30 minutes on, 30 minutes sleep. No pressure.
Program 2:
30 minutes on, 30 minutes sleep. Low pressure.
Program 3:
30 minutes on, 30 minutes sleep. Medium pressure.
Program 4:
30 minutes on, 60 minutes sleep. No pressure.
Program 5:
30 minutes on, 60 minutes sleep. Low pressure.
Program 6:
30 minutes on, 60 minutes sleep. Medium pressure.
Program d:
Drain Mode. Please refer to page 2 in this manual for
details.
10
You can choose to operate in either Manual Mode or Program Mode.
Manual Mode allows you to adjust time and pressure settings as you
choose. Program Mode allows you to choose one of six programs that
provide treatment for a set time then sleep (providing no treatment) for a set time,
continuously, at a specific pressure setting (refer to list of available programs below).
The unit automatically starts in Manual Mode.
COntrOL unit
The exterior of the Control Unit and the visible interior surfaces of the reservoir can be
cleaned with a soft cloth and one of the following cleaning agents:
•
Mild detergent
•
70% Isopropyl alcohol
•
100% Simple Green
®
•
Quaternary ammonium (such as Cavicide
®
or Virex
®
– typically only found in a clinical use
setting)
procedure
:
•
Follow the manufacturer’s instructions and precautions for the cleaning agent you select.
•
Apply the selected cleaner to a soft cloth and wipe down all surfaces of the Control Unit.
•
Allow the Control Unit to dry thoroughly before storing it in the bag.
•
The Control Unit should be cleaned as needed.
dO nOt use:
•
Phenolic-based disinfectants (such as Amphyl
®
– typically only found in a clinical use
setting).
•
Any solvent-based cleaners on the Control Unit. Doing so will damage the plastics and
will void your warranty.
•
Abrasive materials to clean the Control Unit. Doing so will damage the plastics and will
void your warranty.
Caution
: The Control Unit is not a waterproof device. Do not apply a direct stream of any
liquid onto the Control Unit, submerge the Control Unit, or allow any liquid to pool on the
surface of the front panel of the Control Unit.
COnneCtOr HOse
The surface of the Connector Hose can be cleaned using a soft cloth and one of the
following:
•
Mild detergent
•
Steri-Fab
®
•
70% Isopropyl alcohol
•
We do not recommend the use of quaternary ammonium (such as Virex
®
or Cavicide
®
).
dO nOt use
:
•
Phenolic-based disinfectants (such as Amphyl
®
– typically only found in a clinical use
setting).
•
Any solvent based cleaners. Doing so will damage the plastics and will void your
warranty.
•
Abrasive materials. Doing so will damage the plastics and will void your warranty.
•
Any petroleum based lubricants. Doing so will damage the o-rings and will void your
warranty. If lubrication is required, the use of silicon spray is recommended.
WrAPs
Gently remove Heat Exchanger from the Sleeve. Hand wash the Sleeve in cold water,
using a mild detergent or antibacterial soap. Do not use fabric softener. Hang to dry.
If needed, the external surface of the Heat Exchanger may be cleaned by wiping down with
commercial non-bleach cleaning wipes or hand washed using a very small amount of mild
detergent or antibacterial soap. DO NOT MACHINE WASH.
Refer to the Wrap Use Guide accompanying individual Wraps for more information.
s t o r A G e
C l e A N i N G
When you are done using the System for the day:
•
Unplug the AC Adapter and the Connector Hose from the Control Unit.
•
Push the door release button to open the door.
•
Carefully pour out the ice and water.
•
Wipe off any excess water.
•
Store the Control Unit with the lid fully open to allow the interior to dry and preserve
the reservoir seal. Store the Control Unit in the Carry Bag or in another safe location.
Remember that your GRPro Control Unit is a valuable piece of equipment and should
be treated with great care, like a laptop computer.
Storage Temperature Conditions: 33° - 120°F (1° - 50°C)
Relative Humidity: 10% - 95% non-condensing
Caution
: Do not keep in extreme cold or hot temperatures (below 33°F or above 120°F
or below 1° or above 50°C). Do not leave in a hot or freezing car. Do not leave the Control
Unit in direct sunlight. The UV light may damage or discolor the Control Unit.
EN-8
EN-9
SUPERSEDED