background image

Do not use this device without your physician’s specific recommendations for the frequency 

and duration of your treatments.

The most common recommendation is to use cold therapy at least 4 times a day, for approximately 

30 minutes each time, with at least a 30-minute break between treatments. While the 

temperature is adjustable, it is reported that the greatest benefit from cold therapy is in the 

40-60°F (

4.5-15.5°C)

 temperature range. Generally reported compression recommendations 

are for “None” to “Low” pressure settings during the first 24-48 hours after surgery or 

injury, increasing to “Medium” or “High” only if it is comfortable after the first 48 hours. 

We recommend that “High” pressure is never applied if you are confined to bed. 

Be sure to read further safety warnings on pages 12-13 of this User’s Manual.

Modes of oPerAtioN

Manual Mode: 

The System automatically starts in this mode, and allows the user to 

adjust treatment time and pressure settings.

Program Mode:

 This mode allows the user to choose one of six treatment programs that 

provide therapy for a set time then sleep (no treatment) for a set time, continuously, at a 

specific pressure setting.

drain Mode: 

Allows a user to connect a Hose with a special Hose Adapter (purchased 

separately) to the unit, enter Drain Mode and have the unit empty the water out of the ice 

box through the Hose. Drain Mode can be accessed by pressing the program key and 

toggling through all six programs. To empty water in the System while in Drain Mode, press 

the program key until you reach Drain Mode, attach the Hose Adapter to the Connector Hose, 

place the Hose Adapter over a sink, and press the Play button. Drain Mode 

will run the Control Unit’s fluid pump for up to six minutes (long enough 

to fully empty the ice box). Drain Mode is indicated by the following icons:

d e t A i l e d   i N s t r u C t i o N s   f o r   u s e

buttoNs 

Power:

 Use this button to turn the Control Unit on and off.

Program:

 Use this button to select one of the available Programs or to return to 

Manual Mode. See page 8 in this manual for more information on Programs.

Play/Pause: 

Use this button to start or pause a treatment.

Add time: 

Use this button to add time in Manual Mode (does not work in Program 

Mode). You can add up to 90 minutes. Treatment must be paused in order to add  

or subtract time.

subtract time:

 Use this button to reduce time in Manual Mode (does not work in 

Program Mode). You can subtract up to 90 minutes. Treatment must be paused in 

order to add or subtract time.

Pressure selection: 

Use this button to select one of four pressure settings: No 

Pressure, Low Pressure (5-15 mmHg), Medium Pressure (5-50 mmHg), and High 

Pressure (5-75 mmHg). Pressure selection is not available in Program Mode. 

Treatment must be paused to change pressure settings.

Volume: 

Use this button to select the option of sound or no sound. Push to mute 

sound. Alarms will still sound even with Volume off.

C/f button:

 Use this button to select either Celsius or Fahrenheit on the 

temperature display.

backlight:

 Use this button to turn the backlight on or off.

Adjusting temperAture

To adjust the temperature being applied during treatment, simply turn the temperature 
knob towards 3 snowflakes for the maximum amount of cold, or towards 1 snowflake for 
the least amount of cold. Notice that as you adjust the knob, the “Target Temperature” on 
the display will change. The System will automatically adjust to match the temperature you 
have selected. 

tiP

: The maximum amount of cold is dictated by the amount of ice in the reservoir and the 

setting of the temperature adjustment knob. You may need to stir or replenish the ice to 
achieve coldest temperatures. Ice will melt faster in the acute phases of injury as the Game 
Ready System is removing greater amounts of heat from the treatment site.

disPlAY

status bar:

Off (Mute)

On

Indicates selected mode.

Indicates unit is paused.

Indicates unit is running.

other icons:

Indicates an error. See troubleshooting 
pages 15-17 for error codes.

Indicates sleep mode time.

Indicates remaining battery capacity.

temperature:

              

Actual temperature: 

Approximate 

temperature of the water leaving the Control 
Unit.

        

target temperature: 

Indicates the target 

temperature setting. The Control Unit will 

automatically match the Actual Temperature 

and Target Temperature as closely as possible.

Error

Battery

Sleep

EN-2

EN-3

SUPERSEDED

Summary of Contents for GRPro 2.1

Page 1: ...odel Numbers 550550 03 550550 03 RN 550550 53 GRPro 2 1 Control Unit User s Manual53 Game Ready Control Unit User s Manual PN 703758 Rev D 2011 CoolSystems Inc All rights reserved Brought to you by Co...

Page 2: ...of Operation______________________________________ 2 Buttons_________________________________________________ 2 Adjusting Temperature___________________________________ 3 Display______________________...

Page 3: ...n on Programs Play Pause Use this button to start or pause a treatment Add Time Use this button to add time in Manual Mode does not work in Program Mode You can add up to 90 minutes Treatment must be...

Page 4: ...a table during use Note that using the System in an environment with a high ambient temperature may affect its ability to provide adequate cooling or may limit the ice life If you will be using the S...

Page 5: ...Make sure you hear it click To disconnect simply press the blue or gray button and remove the connector from the Wrap 9 Set your target temperature displayed in the Target Temperature window by turnin...

Page 6: ...id your warranty Abrasive materials to clean the Control Unit Doing so will damage the plastics and will void your warranty Caution The Control Unit is not a waterproof device Do not apply a direct st...

Page 7: ...tomer Service at 1 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor 1 Identify the filter within the ice reservoir 2 Using two fingers grasp and squeeze the tw...

Page 8: ...e in the acute stages of inflammatory phlebitis in the affected region Who have any history or risk factors for deep vein thrombosis or pulmonary embolus including prolonged bed rest in the affected r...

Page 9: ...armful ingress of water 5 3 This product provides ordinary protection against ingress of water Degree of safety in the presence of flammable anesthetics or oxygen 5 5 Not suitable for use in an oxygen...

Page 10: ...ufacturer s Declaration Electromagnetic Immunity The GRPro 2 1 is intended for use in the electromagnetic environment specified below The customer or the user of the GRPro 2 1 should assure that it is...

Page 11: ...g NO Pressure wrap shall be vented to atmosphere The cooling temperature of the circulating ice water will be adjustable between 34 F 1 C and 50 F 10 C as long as the ice water in the ice box is suppl...

Page 12: ...e connections Disconnect and reconnect Wrap from Connector Hose Make sure there is water in the ice box Verify that the ice box filter is not clogged Make sure there are no kinks in the Wrap or Connec...

Reviews: