Do not use any external power supply (mobile phone charger) other than the one
provided by your Healthcare Professional who prescribed you the medical device.
Always refer to the device manual for proper charging instructions.
●
Use of accessories and cables other than those specified or provided by the
manufacturer of this medical device could result in improper operation of the device.
Always refer to the manufacturer indications.
●
Operate the Gait Up GO medical device at ambient room conditions in terms of
temperature (0°C - 30°C), humidity (10%-70% - non-condensing) and pressure (600 to
1070 hPa).
●
Do not introduce any pointy object in the hole present on the backside of the Physilog6
sensor device accessory. On the internal face of the hole, a membrane that ensures the
IP64 protection index of the sensor device is placed. By damaging it, the protection is
no longer ensured and water ingress can damage the electronics and battery housed in
the enclosure of the Physilog6. Be aware that there is no reset button behind such hole
in the device.
1.1.2. Warnings related to the conditions and awareness of the end-users
●
DO NOT use the Gait Up GO in other parts of the Participant body but in the ones
indicated in the Instructions For Use.
●
DO NOT manipulate or remove Gait Up Go accessories from the user body during the
session. Accurate results are only guaranteed if the Physilog6 sensor accessory is firmly
attached on the user shoe or belt and the Instructions for Use are followed during the
duration of the assessment.
●
Keep device away from pets
1.2. Cautions
●
Operate Gait Up GO medical device only as prescribed by your Healthcare Professional.
Precaution should be exercised when using the device in following conditions:
●
Do not use Gait Up GO medical device in case of software malfunctioning, including
mobile phone application, repeated Bluetooth Low Energy (BLE) connection issues,
and repeated reset of the Physilog6 sensor device accessory. In all those cases, contact
your Healthacre Professional who has prescribed you the medical device.
●
Gait Up GO medical device uses a single Physilog6 sensor device accessory. A
Participant must not wear more than one Physilog6 accessory on the body at the same