7. Measurement
Turn ON the power and the measurement starts.
Starting Screen
When the power is turned ON, the channel number configured on the LX-
7230N is displayed at the top of the LCD.
Make sure whether the channel number on the
LCD matches the channel number indicated on
the label of the LX-7230N and the channel
number configured on the receiving monitor.
This screen automatically moves onto the next
waveform display screen.
Waveform Display Screen
ECG waveform (1CH when using 3-electrode lead cable, 2CH when using
other lead cable), heart rate, pacemaker marker, respirogram, respiration rate,
pulse wave, SpO
2
measurement value, remaining battery level, and various
messages are displayed.
C A U T I O N
The LX-7230N does not have a diagnostic function. Check the
diagnostic function on the receiving monitor.
The LX-7230N does not have an alarm function. Check the alarm
function on the receiving monitor.
The ECG waveform size and sweep speed settings displayed on the
LCD of the LX-7230N do not interface with the ones displayed on the
screen of the receiving monitor.
The heart rate and respiration rate displayed on the LCD of the LX-
7230N may be different from the ones displayed on the receiving
monitor. Because the algorithm of the ECG and respiration rate is
different.
Summary of Contents for LX-7230N
Page 22: ...1 General Description 2 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 66: ...7 Measurement 46 Blank Page ...
Page 72: ...8 Operation 52 Blank Page ...
Page 78: ...10 Changing the Transmitter Channel and Group ID 58 Blank Page ...
Page 84: ...11 Troubleshooting 64 Blank Page ...
Page 92: ...14 Standard and Optional Accessories 72 Blank Page ...