4. ECG Monitoring
When the transmitter is used without the SpO
2
probe (sensor), it will measure
only ECG and respiration.
C A U T I O N
When using the transmitter with only the ECG lead cable, SpO
2
measurements on the receiving monitor shall be turned off to prevent an
erroneous alarm.
Connecting the ECG Lead Cable and Electrodes
The optional ECG lead cables for LX-7230N are as follows.
ECG Lead Cables
AHA color code:
Item No.
Applicable Lead
Remark
CMT-01HTH-0.8DA
Limb Lead (1CH)
3-electrode Hook Type
(White, Black, Red)
CMT-02HTH-0.8DA
Limb Lead (2CH)
4-electrode Hook Type
(White, Black, Red, Green)
CMT-03HTH-0.8DA
Limb Lead (1CH)
+Chest (1CH)
5-electrode Hook Type
(White, Black, Red, Green, Brown)
CMT-01FTH-0.8DA
Limb Lead (1CH)
3-electrode Clip Type
(White, Black, Red)
CMT-02FTH-0.8DA
Limb Lead (2CH)
4-electrode Clip Type
(White, Black, Red, Green)
CMT-03FTH-0.8DA
Limb Lead (1CH)
+Chest (1CH)
5-electrode Clip Type
(White, Black, Red, Green, Brown)
W A R N I N G
Use only the specified lead cable from Fukuda Denshi. Otherwise, proper
monitoring may not be performed, and also it may fail defibrillation or cause
a malfunction of the equipment when the equipment is used with a
defibrillator.
Summary of Contents for LX-7230N
Page 22: ...1 General Description 2 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 66: ...7 Measurement 46 Blank Page ...
Page 72: ...8 Operation 52 Blank Page ...
Page 78: ...10 Changing the Transmitter Channel and Group ID 58 Blank Page ...
Page 84: ...11 Troubleshooting 64 Blank Page ...
Page 92: ...14 Standard and Optional Accessories 72 Blank Page ...