15. Specification
75
Wavelength:
Approx. 660nm (Red light)
Approx. 900nm (Infrared light)
Optical output power: Under 15mW
Measurement Response Time:
6 to 7 sec. (Averaging time)
Measurement Value Update Rate: 1 second
N O T E
The SpO
2
measurement accuracy is determined based on the values of
the root-mean-square (rms) difference between SpO
2
readings of the pulse
oximeter equipment and values of SaO
2
determined with a CO-oximeter,
by healthy adult volunteers. The pulse oximeter equipment measurements
are statistically distributed; ±2% measurement accuracy means that only
about two-thirds of pulse oximeter equipment measurements can be
expected to fall within ±2% of the value measured by a CO-oximeter.
US Patents
Covered by one or more of the following U.S. patents and foreign
equivalents:
5,485,847; 5,676,141; 5,743,263; 6,035,223; 6,226,539; 6,411,833;
6,463,310; 6,591,123; 6,708,049; 7,016,715; 7,039,538; 7,120,479;
7,120,480; 7,142,142; 7,162,288; 7,190,985; 7,194,293, 7,209,774;
7,212,847; 7,400,919.
Pulse Wave
Pulse Rate Measurement Range:
20 to 250bpm
Measurement Accuracy:
±3bpm
Measurement Response Time:
6 to 7 sec. (Averaging time)
Measurement Value Update Rate:
1 second
Transmission Method
Modulation Mode:
Digital, Frequency shift keying
Frequency:
1395 to 1400MHz , 1427 to 1432 MHz
Oscillation Method:
PLL Synthesizer method by crystal control
Channel Spacing:
25kHz
Occupied Frequency
Bandwidth:
16kHz
RF output power:
5mW ±2dB
Transmitting Antenna:
ECG lead cable and/or SpO
2
Probe
Summary of Contents for LX-7230N
Page 22: ...1 General Description 2 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 66: ...7 Measurement 46 Blank Page ...
Page 72: ...8 Operation 52 Blank Page ...
Page 78: ...10 Changing the Transmitter Channel and Group ID 58 Blank Page ...
Page 84: ...11 Troubleshooting 64 Blank Page ...
Page 92: ...14 Standard and Optional Accessories 72 Blank Page ...