3. Preparation
1) Installing the Batteries
The LX-7230N functions with two "AA" size (“LR06” size) alkaline batteries.
With new batteries, the LX-7230N is capable of the following operation.
LX-7230N: approximately 2.5 days
(However, continuous operating time may be shorter than the above
mentioned time depending on the application of the SpO
2
probe (sensor).)
W A R N I N G
Unplug the ECG lead cable when the battery compartment lid is
opened. Otherwise, patient leakage current beyond the allowable value
may occur.
Use new "AA" size (“LR06” size) alkaline batteries.
Do not short out the (+) and (-) terminals. It may result in exothermic
heat and fire, the leakage from the batteries may damage the
equipment, or an explosion from the batteries may occur.
Install the batteries with the correct polarity.
Do not use a disassembled or a damaged battery due to drop or shock.
The leakage from the batteries may damage the equipment, or an
explosion from the batteries may occur.
Do not use different types of batteries at the same time. The leakage
from the batteries may damage the equipment, or an explosion from the
batteries may occur.
Remove the exhausted batteries immediately. The leakage from the
batteries may damage the equipment, or an explosion from the batteries
may occur.
If the transmitter is not in use for a long period of time, remove the
batteries and store the equipment in an appropriate place. If the
batteries are left in the transmitter for a long period of time, the leakage
from the batteries may damage the equipment or an explosion from the
batteries may occur.
Make sure to replace the two batteries simultaneously. If new and used
batteries are mixed, a leakage from the batteries may damage the
equipment or an explosion from the batteries may occur.
C A U T I O N
Use only alkaline batteries. Other batteries will shorten the continuous
operating time.
Once the power switch is on the OFF position, then open the battery
compartment lid.
Do not replace the batteries with wet hands.
In case of storing the used or unused batteries, make sure that the
terminals are not touching other batteries or metal parts.
Summary of Contents for LX-7230N
Page 22: ...1 General Description 2 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 66: ...7 Measurement 46 Blank Page ...
Page 72: ...8 Operation 52 Blank Page ...
Page 78: ...10 Changing the Transmitter Channel and Group ID 58 Blank Page ...
Page 84: ...11 Troubleshooting 64 Blank Page ...
Page 92: ...14 Standard and Optional Accessories 72 Blank Page ...