Maintenance
3 - 27
5.Counter electrode voltage
(1)
Check that the change in the recorded sensitivity is within
±
5%
when
±
550 mV of DC voltage is imposed to the input of the
Electrocardiograph.
(2)
Check that the change in the recorded sensitivity is within
±
5%
when
±
550 mV of DC voltage is imposed between the two input
terminals of the Electrocardiograph and the RF(N) electrode fitting.
6.Internal noise
Make shortcircuited the each terminal of all input terminals with the
parallel impedance with 51k
Ω
and 47nF. Place the baseline at the center
at each lead selection position, set the sensitivity to 2, and record the noise.
Check that the size of the noise on the recorded line is 0.4 mmp-p or
smaller at random 0.2-second intervals.
7.Recording speed
Connect a clock that provides the unit with signal voltage at 1-second
intervals, and record the pulse waveform. Check that the error in transfer
speed at speeds of 5mm/sec, 10 mm/sec, 12.5mm/sec, 25 mm/sec, and 50
mm/sec is within
±
3%.
Summary of Contents for FCP-7101
Page 10: ...viii ...
Page 12: ......
Page 18: ...Chapter 1 1 6 ...
Page 20: ......
Page 32: ...Chapter 2 2 12 ...
Page 66: ...Chapter 3 3 32 ...
Page 68: ......
Page 76: ...5 2 ...
Page 78: ...5 4 ...
Page 124: ......
Page 136: ...Chapter6 6 12 ...
Page 138: ......
Page 139: ...7 1 MAIN BOARD PCB 6950 component side ...
Page 140: ...7 2 MAIN BOARD PCB 6950 soldering side ...
Page 141: ...7 3 KEY BOARD PCB 6949 component side ...
Page 142: ...7 4 KEY BOARD PCB 6949 soldering side ...
Page 143: ...7 5 SENSOR BOARD PCB 6948 component side ...
Page 144: ...7 6 MAG UP BOARD PCB 6951 component side ...
Page 146: ......
Page 155: ...8 9 Outline view FX 7102 ...
Page 156: ...8 10 Outline view FCP 7101 ...
Page 160: ...8 14 ...
Page 162: ......
Page 169: ......