Chapter 2
2 - 10
●
Recording paper is fed, but printing and waveform recording are not
performed.
Item to Check
Method of Checking and Corrective Action
Place to Check
Connector
Verify that the thermal print head cable is securely
attached to connector P3 on the MAIN BOARD.
Verify also that the soldering of connectors P3 is free
of defect.
Verify that the thermal print head and thermal print
head cable are securely connected and that the
thermal print head cable is not broken.
P3
Thermal print head cable
REFERENCE
P5-33
Power supply
Verify that pins No. 1
,
2
,
13 and 14 of P3 on the
POWER BOARD are supplied with +24 volts during
recording.
P3 pins 1
,
2
,
13and 14
REFERENCE
P5-33
Control signal
When recording, verify that the signals of
[LATCH*] (pin No. 10), [CLOCK] (pin No. 9), and
[STROBE1*-4*] (pins No. 7and8) of P3 on the
MAIN BOARD are not fixed at "H" or "L."
If the above is abnormal, check the signals for IC7B
to IC8, 56-59.
P3 pins 7 to 10
IC8pins 56 to59
REFERENCE
P5-29,33
Overheat
Printing is not performed if the temperature of the
thermal head is 60
°
C or higher.
Check that IC15 pin 7 is not at “L” level.
Check the maker maintenance status to make sure
that the temperature of the thermal head is not set to
60
°
C or higher.
IC15 pin 7
Maker maintenance status
REFERENCE
P5-33, P3-14
●
Recording paper is not fed
Item to Check
Method of Checking and Corrective Action
Place to Check
Connector
Check that the motor cable is securely connected and
that it is not broken.
P5
Motor cable
REFERENCE
P5-34
Control signal
While recording, check that a signal is present at P5
pins 1 through 4 on the MAIN BOARD.
Also check that the level is not "L" at Q6 and Q7
pins 3.
If a signal is present, the motor may be faulty.
P5
REFERENCE
P5-34
●
A certain lead record appears as a baseline.
Item to Check
Method of Checking and Corrective Action
Place to Check
Patient cable
Check the patient cable. for a break and check the
connection staus.
Patient cable
OP Amp of ECG
AMP, A/D converter
and limiter
Check that the operation of the OP Amp, IC23 to
IC27, and A/D converter IC37 (which are related to
the baseline lead) and the transistors Q13 to Q21 for
the protection against excess input are not shorted.
IC23 to IC27, IC37,Q13 to
Q21
REFERENCE
P5-35,36
Summary of Contents for FCP-7101
Page 10: ...viii ...
Page 12: ......
Page 18: ...Chapter 1 1 6 ...
Page 20: ......
Page 32: ...Chapter 2 2 12 ...
Page 66: ...Chapter 3 3 32 ...
Page 68: ......
Page 76: ...5 2 ...
Page 78: ...5 4 ...
Page 124: ......
Page 136: ...Chapter6 6 12 ...
Page 138: ......
Page 139: ...7 1 MAIN BOARD PCB 6950 component side ...
Page 140: ...7 2 MAIN BOARD PCB 6950 soldering side ...
Page 141: ...7 3 KEY BOARD PCB 6949 component side ...
Page 142: ...7 4 KEY BOARD PCB 6949 soldering side ...
Page 143: ...7 5 SENSOR BOARD PCB 6948 component side ...
Page 144: ...7 6 MAG UP BOARD PCB 6951 component side ...
Page 146: ......
Page 155: ...8 9 Outline view FX 7102 ...
Page 156: ...8 10 Outline view FCP 7101 ...
Page 160: ...8 14 ...
Page 162: ......
Page 169: ......