Chapter 3
3 - 22
Update
The software installed in the ECG can be updated following the procedure shown below.
1. Select [PROGRAM UPDATE] on the [MAKER MAINTENANCE] screen, and
the following screen is displayed.
2. Enter the IP address of the ECG and that of the HOST, using the [
←
], [
→
], [
↑
],
or the [
↓
] key.
3. Press the [ENTER] key, and then select software with the [ID] key according to
the destination by referring to the following table.
Name of software file to be updated
Destination
“FX7102EN.BIN” General
export
“FX7102CN.BIN” China
“FX7102JP.BIN” Within
Japan
4. Connect the LAN cable to the LAN port and push the [START/STOP] key, and
the UPDATE is started.
To update the software, the environment where LAN can be connected
is required.
On completion of updating, the ECG is automatically reset.
Note
Note
Summary of Contents for FCP-7101
Page 10: ...viii ...
Page 12: ......
Page 18: ...Chapter 1 1 6 ...
Page 20: ......
Page 32: ...Chapter 2 2 12 ...
Page 66: ...Chapter 3 3 32 ...
Page 68: ......
Page 76: ...5 2 ...
Page 78: ...5 4 ...
Page 124: ......
Page 136: ...Chapter6 6 12 ...
Page 138: ......
Page 139: ...7 1 MAIN BOARD PCB 6950 component side ...
Page 140: ...7 2 MAIN BOARD PCB 6950 soldering side ...
Page 141: ...7 3 KEY BOARD PCB 6949 component side ...
Page 142: ...7 4 KEY BOARD PCB 6949 soldering side ...
Page 143: ...7 5 SENSOR BOARD PCB 6948 component side ...
Page 144: ...7 6 MAG UP BOARD PCB 6951 component side ...
Page 146: ......
Page 155: ...8 9 Outline view FX 7102 ...
Page 156: ...8 10 Outline view FCP 7101 ...
Page 160: ...8 14 ...
Page 162: ......
Page 169: ......