Technical Instructions Related to EMC
9 - 3
Immunity information (to other than radiation/conduction)
The following is a declaration related to the electromagnetic immunity of
the equipment.
Guidance and declaration by manufacturer-Electromagnetic emission-CISPR14
The equipment is intended to be used in the electromagnetic environment designated below.
Client or user of the equipment is requested to confirm the equipment will be operated in such Environment.
Immunity test
IEC 60601
test level
Conformity level
Guidance for electromagnetic
environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6kV in contact.
±
8kV in air.
±
6kV in contact.
±
8kV in air.
The floor must be wood, concrete or
ceramic tiles. If the floor is covered
with synthetic material, the relative
humidity must be at least 30%.
Electrical first
transient/burst
IEC 61000-4-4
±
2kV against
power supply
line.,
±
1kV against
input/output
line.
±
2kV against
power supply
Line.
±
1kV against
input/output
Line.
The quality of main source must be
the same as standard commercial
use or hospital environment.
Surge
IEC 61000-4-5
±
1kV in
differential mode.
±
2kV in common
mode.
±
1kV in
differential mode.
±
2kV in common
mode.
The quality of main source must be
the same as standard commercial
use or hospital environment.
<5%U
T
(drop>95%U
T
)
for 0.5 cycle.
<5%U
T
(drop>95%U
T
)
for 0.5 cycle.
40%U
T
(drop 60%U
T
)
for 5 cycle.
40%U
T
(drop 60%U
T
)
for 5 cycle.
70%U
T
(drop 30%U
T
)
for 25 cycle.
70%U
T
(drop 30%U
T
)
for 25 cycle.
Voltage dip,
short-time failure
and voltage
fluctuation on
power supply input
line
IEC 61000-4-11
<5%U
T
(drop >95%U
T
)
for 5 seconds.
<5%U
T
(drop >95%U
T
)
for 5 seconds.
The quality of main source must be
the same as standard commercial
use or hospital environment.
If user of the equipment requires a
continuous operation even during
failure of main source, use of
uninterruptible power supply or
battery is recommended as power
source for the equipment.
Magnetic field
intensity of source
frequency
(50/60Hz)
IEC 61000-4-8
3 A/m
3 A/m
The magnetic field intensity of
source frequency must be
comparable to the level
characteristics of the magnetic field
in standard commercial or hospital
environment.
Note: As U
T
,
AC main source voltage has the precedence of applied test level.
Summary of Contents for FCP-7101
Page 10: ...viii ...
Page 12: ......
Page 18: ...Chapter 1 1 6 ...
Page 20: ......
Page 32: ...Chapter 2 2 12 ...
Page 66: ...Chapter 3 3 32 ...
Page 68: ......
Page 76: ...5 2 ...
Page 78: ...5 4 ...
Page 124: ......
Page 136: ...Chapter6 6 12 ...
Page 138: ......
Page 139: ...7 1 MAIN BOARD PCB 6950 component side ...
Page 140: ...7 2 MAIN BOARD PCB 6950 soldering side ...
Page 141: ...7 3 KEY BOARD PCB 6949 component side ...
Page 142: ...7 4 KEY BOARD PCB 6949 soldering side ...
Page 143: ...7 5 SENSOR BOARD PCB 6948 component side ...
Page 144: ...7 6 MAG UP BOARD PCB 6951 component side ...
Page 146: ......
Page 155: ...8 9 Outline view FX 7102 ...
Page 156: ...8 10 Outline view FCP 7101 ...
Page 160: ...8 14 ...
Page 162: ......
Page 169: ......