12
Clean and disinfect system and transducer to a high level (semi-critical uses)
b
Ensure that the connector and most of the cable remain out of the fluid. You may submerge 2 inches
(5 cm) of the cable proximal to the transducer scanhead.
Figure 1
Transducer component names
12 Rinse
the transducer
three separate times
using the following procedure:
a
Rinse the transducer in clean, running water according to the disinfectant manufacturer’s instructions
(at least one minute).
b
Ensure that the connector and at least 12-18 inches (31-46 cm) of the cable from the connector remain
dry.
c
Repeat this step until the transducer has been rinsed three separate times.
13 Dry
the transducer with a sterile lint-free cloth.
14 Dispose
of the disinfectant according to the manufacturer’s guidelines.
Caution
Do not soak the transducer longer than recommended by the chemical
manufacturer.
Do not immerse the transducer connector in any disinfectant solution.
Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a
non- recommended disinfecting solution or incorrect solution strength can
damage or discolor the transducer and void the warranty.
WARNING
High-level disinfectants can cause harm to the patient if not completely removed
from the transducer. Follow the manufacturer’s rinse instructions to remove
chemical residue.
Maximum
5 cm
(2 inches)
Do not submerge
Scan head
Cable
Connector
Summary of Contents for SonoSite Edge II
Page 1: ...User Guide Errata ...
Page 40: ...38 Umgebungsbedingungen ...
Page 132: ...130 Omgevingslimieten ...
Page 206: ...204 Περιβαλλοντικοί περιορισμοί ...
Page 226: ...224 Предельные значения параметров окружающей среды ...
Page 261: ...Dansk Norsk Svenska Ελληνικά Русский Türkçe 繁體中文 環境限制 259 ...
Page 262: ...260 環境限制 ...
Page 263: ......
Page 264: ...P19439 02 P19439 02 ...