Clean and disinfect system and transducer to a high level (semi-critical uses)
11
English
De
ut
sch
Español
Fr
a
n
ç
a
is
Ita
lia
no
P
o
rtuguês
Nederlands
8 Verify
that all gel, debris, and bodily fluids have been removed from the system and transducer. If
necessary, repeat steps 5, 6 & 7 with a new wipe.
9
Examine
the system, transducer and cable for damage, such as cracks or splitting where fluid can enter.
If damage is evident, do not use. Instead, contact FUJIFILM SonoSite or your local representative.
10 Prepare
the disinfectant for use.
a
Choose a high-level disinfectant from the list of approved disinfectants.
b
Check the expiration date on the bottle to ensure the disinfectant has not expired.
c
Mix or check that the disinfection chemicals have the concentration recommended by the manufacturer
(for example, a chemical strip test).
d
Check that the temperature of the disinfectant is within the manufacturer’s recommended limits.
11 Perform
a high-level disinfection of the transducer. Use the following procedure:
a
Immerse the transducer in a high-level disinfectant solution.
WARNING
Failure to remove all gel, debris, and bodily fluid could leave contaminants on the
probe.
High-level disinfectants compatible with SonoSite Edge II transducers
Disinfectant
Compatible transducers
Temperature
Disinfectant Soak Duration
Cidex
C8x, C11x, C35x, HFL38xi,
HFL50x, HSL25x, ICTx, L25x,
L38xi, P10x, P11x
1
, rC60xi,
rP19x
25°C, 77°F
45 minutes
Cidex OPA
C8x, C11x, HFL38xi, HFL50x,
HSL25x, ICTx, L25x, L38xi,
L52x
2
, P10x, rC60xi, rP19x
20°C, 68°F
12 minutes
1
The P11x Transducer is not licensed for use in Canada.
2
The L52x Transducer is for veterinary use only.
Refer to the cleaners and disinfection document available at
for a complete list of the most
current cleaners and disinfectants.
WARNING
High-level disinfectants can cause harm to the patient if not completely removed
from the transducer. Follow the manufacturer’s rinse instructions to remove
chemical residue.
Summary of Contents for SonoSite Edge II
Page 1: ...User Guide Errata ...
Page 40: ...38 Umgebungsbedingungen ...
Page 132: ...130 Omgevingslimieten ...
Page 206: ...204 Περιβαλλοντικοί περιορισμοί ...
Page 226: ...224 Предельные значения параметров окружающей среды ...
Page 261: ...Dansk Norsk Svenska Ελληνικά Русский Türkçe 繁體中文 環境限制 259 ...
Page 262: ...260 環境限制 ...
Page 263: ......
Page 264: ...P19439 02 P19439 02 ...