-4 Calibration gas flow time setting
Press the FUNC key at the calibration cycle indication
and “
” will appear.
The AUTO CAL LED of the function indicating lamps
will flash.
Press the
key and the auto calibration gas flow time
is settable.
Set the flow time on the sub indication.
The numeric value is incremented by pressing the
key.
Press the
key and the digit can be selected.
The setting range is 100 to 599 seconds.
Press the ENT key when the calibration gas flow time
has been set. The data is stored in memory.
The times set in this mode is common to all the cali-
bration gas flow time including replacement time.
Note: When the auto calibration start time has been set, set the auto calibration gas flow mode.