26
NO
NO
CAL
ENT
∧
Press the CAL key. “
” appears in the main indica-
tion and “ “ appears in the sub indication.
At each press of the
key, the setting is changed in the
order of “
→ → →
”.
After selecting, press the ENT key. The data is stored in
memory.
= Meaning of set value =
Select
Zero calibration
code
Manual calibration
Auto calibration
0
Calibration of all components
Calibration of all components
1
Calibration of components other than O
2
Calibration of components other than O
2
2
Calibration of O
2
Note 1: This mode is invalid when O
2
component is not included.
Note 2: The main indication flickers when zero calibration
( ZERO + CAL ) is made on selected components.
Setting of single calibration of all or each range:
This setting operation is used for calibration of all or each range during zero/span calibration.
Carry out the setting as follows.
By pressing CAL key while the highest digit of the 1st
components main indication is flashing, “
” will
appear on the main indication and “ ” will appear on the
sub indication.
Press
key and “ ” will change to
“ ”.
= Meaning of set values =
: The calibration is valid only for the selected range,
and zero and span calibration can be made indepen-
dently for each range.
: By conducting zero and span calibration for one
range, the calibration will be done automatically for
the other ranges as well.
Note 3: S.CAL “1” is interlocked with the range ratio
calibrated by “0” on each range.
Note 4: When a 1-range meter is used, be sure to set
S.CAL to “0”. If “1” is set, the result is the
same as “0”.
Press the ENT key when the selection is finished.
Note: When using air for the high range and 9-10 vol % O
2
/N
2
for the low range, be sure to set the
value of concentration for each range.
NO
NO
CAL
∧
ENT
NO
NO