-3 Auto calibration cycle setting
The calibration cycle is settable in either ‘days’ or ‘hours’.
Press the FUNC key at the calibration start time indica-
tion and “
” will appear.
The AUTO CAL LED of the function indicating lamps
will flash.
Press the
key and the auto calibration cycle is
settable. The highest digit on the sub indication will flash.
Select either ‘days’ or ‘hours’ by pressing the
key.
“ ” appears on the sub indication when selecting
‘days’ and “ ” appears when selecting ‘hours’.
Press the
key and the calibration cycle can be set in
the lower digits of the sub indication.
The numeric value is incremented by pressing the
key.
Press the
key and the digit can be selected.
Day is settable in a range of 1 to 7 days. Hour is
settable in a range of 1 to 99 hours.
If “
” (hours) is set, it will be automatically set to
“ ” (hour).
Press the ENT key when the calibration cycle has been
set. The data is stored in memory.
Note:When the auto calibration start time has been set, then set the auto calibration gas flow time.