31
NO
NO
NO
NO
FUNC
RMT RANGE
>
ENT
∧
t
t
t
t
t
Remote range setting (option)
This is used for selecting the range via an external signal.
This setting should be OFF when remote range is not
used.
Press the FUNC key in the measuring status and
“
” will be indicated.
The RMT RANGE LED of the function indicating lamp
will flash.
Press the
key and the remote range function is
settable.
The sub indication will flash.
Select the remote range “
” or “
” by pressing the
key.
The setting is valid for all the components.
Press the ENT key when the remote range setting is
finished. Setting operation is now completed.
When the remote range function is set ON, the RMT
RANGE LED of the function indicating lamps will
remain lit, and the keys will be unusable for range
selection.
• By inputting 5V DC between the remote input
terminals
⑬
and
⑭
, the No. 1 range is selected.
• No. 2 range is selected when 5V DC input is not
applied to the remote input terminals
⑬
and
⑭
.