output Hold
By setting an output hold to ON, an output signal of each channel are held during the manual/auto
calibration and for the gas flow time (refer to Item 6.4, Setting of Auto Calibration). Regardless
of Hold ON/OFF setting, an output signal can be held via an external input.
1 Manual calibration
• Hold setting
: 0 to 100% FS
• Response time
: 1 to 60 sec.
(Initial value: 15 sec)
• Average period
: 1 to 59 min or 1 to 4 hours
(Initial value: 1 hour)
1 to 59 minutes when the unit is set to minute and 1 to 4 hours when it
is set to hour.
• Backlight Timer
: 1 to 60 min (Initial value: 5 min)
• Maintenance mode : 0000 to 9999
(Initial value: 0000)
Setting Range
Calibration
Output hold
Press the key to flow gas.
To execute calibration, press the
key.
Hold extending
time
Time set to gas flow time (See Item 6.4 Auto Calibration.)
or
ENT
ENT
ZERO
SPAN
2 Auto calibration
3 Remote hold
Calibration
Output hold
Hold extending time.
Auto calibration start
Auto calibration end
Time hold set to the gas flow time
4 Screen display during Holding
The “Hold ON” message blinks on the measuring screen.
Since the screen displays the process of calibration during the manual calibration, “Hold ON” is
not displayed even if the output signal is held, but the screen is displayed with the hold extending
time.
Remote hold Input
Output hold
With input
Without input