5
INZ-TN2ZRE-E
3.1 installation conditions
To install the analyzer for optimum performance, select a location that meets the following
conditions;
(1) This instrument is system built in type. This instrument should be used while embedded in a
panel, locker, or enclosure of steel sheet.
(2) Use this instrument indoors.
(3) A vibration-free place
(4) A place which is clean around the analyzer.
(5) Power supply
Rated voltage
: 100V to 240V AC
Operating voltage
: 85V to 264V AC
Rated frequency
: 50/60 Hz
Power consumption
: 100 VA max.
(6) Operation conditions
Ambient temperature : -5° to 45°C (max. 40°C when two optical units are used, and the
power supply is more than 200V AC)
Ambient humidity
: 90 % RH or less, no condensation
Altitude
: Up to 2,187yard [2,000m]
Installation category
: II
Pollution Degree
: 2
(7) A breaker that meets IEC60947-1 and IEC60947-3 should be installed in a building equip-
ment.
(8) A breaker should be installed near an analyzer where an operator can handle it easily.
(9) A label saying that the “Breaker for the analyzer” should be labeled on a breaker.
(10) The breaker rating should meet the analyzer rating and a breaker should be an approved
product.
(11) Do not install the analyzer to the place where it is difficult to handle the power supply cable.
3. instALLAtion
This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent
explosion, fire or other serious accidents.
• Entrust the installation, movement or re-installation to a specialist or the supplier. A poor
installation may cause accidental tipover, electric shock, fire, injury, etc.
• The gas analyzer is heavy. It should be installed with utmost care. Otherwise, it may tipover
or drop, for example, causing accident or injury.
• For lifting the gas analyzer, be sure to wear protective gloves. Bare hands may invite an
injury.
• This unit should be installed in a place which conforms to the conditions noted in the in-
struction manual. Otherwise, it may cause electric shocks, fire or malfunction of the unit.
• During installation work, care should be taken to keep the unit free from entry of cable chips
or other foreign objects. Otherwise, it may cause fire, trouble or malfunction of the unit.
DANGER
CAUTION