34
INZ-TN2ZRE-E
6.3 Alarm setting
6.3.1 setting of alarm values
The High/Low limit alarm output setting for the measured concentration setting can be made.
Arbitrary 5 alarm contact outputs can be used.
To change alarm setting, set the alarm ON/OFF setting to OFF, and then change the value.
( )
ENT
Note
Set the values so that H-limit value > L-limit
value and that (H-limit value
-
L-limit value)
> hysteresis.
When “0” is set, the alarm operation is not
performed.
(1) Enter the “Setting of Alarm No.” screen
from the user mode, and the display
shown at right appears. Point the cur-
sor to the Alarm No. or hysteresis you
want to set by pressing or the key.
Press the
ENT
key.
(2) Select the alarm 1 to 6 to display the
screen shown at right. Operate the
or the key until the cursor is aligned
with a desired item and press the
ENT
key.
( )
ENT
(3) After setting, the alarm setting is now
completed by pressing the
ENT
key.
( )
ENT
End of Alarm setting
To close the “Alarm Setting” or to cancel this
mode midway, press the key.
A previous screen will return.
To close the "Alarm Setting"
ESC
0% to 100% FS (Settable in each range).
Setting range
Cursor for setting value