INZ-TN2ZRE-E
6.5 setting of auto zero calibration
6.5.1 Auto zero calibration
Auto zero calibration is automatically carried out at the time when zero calibration is set.
Components for which a calibration is to be made are determined by setting of auto calibration
component in Item 6.2.4.
Before changing the setting of auto zero calibration, set the ON/OFF to OFF.
( )
ENT
ENT
Description of setting items
• Start Time : Setting at the first calibration
(day of the week, hour, minute)
• Cycle
: A period between the start time of one
calibration and another
(unit : hour/day)
• Flow Time : The time required for the calibration gas
to be replaced in the cell
• ON/OFF : ON/OFF of auto zero calibration
To close the “Setting of Auto Zero Calibration”
or cancel this mode midway, press the key.
A previous screen will return.
To close "setting of Auto Zero Calibration"
ESC
(1) Enter the “Setting of Auto Zero Calibra-
tion” screen from the user mode, and the
display shown at right appears. Operate
the or the key until the cursor is
aligned with a desired item and press the
ENT
key.
(2) In the “Setting of Auto Zero Calibration”
screen that appears, perform the value
entry or the setting. For the value entry
or setting change, use the or the
key and the key to move the cursor to
the right.
After setting, press the
ENT
key, and auto
zero calibration is carried out by the en-
tered setting value.
End of Auto Zero Calibration setting
Press the or
the key, and
date and time are
displayed
alternately.