6.2.4 setting of auto calibration component/range
Select the Ch (component) and the range with which auto calibration is to be performed. The Ch
for which “AR” has been selected as range switch mode is calibrated in the range set here even
when manual calibration is performed.
(1) Select < User mode > → < Calibration
parameters > → < Auto calibration com-
ponent/range >. “Auto Calibration Com-
ponent Range” setting screen appears as
shown at right.
(2) Select the Ch you want to change by
pressing the or the key. Press the
ENT
key and the selected cursor is high-
lighted.
(3) Select the range to be calibrated mainly
by pressing the or the key.
(4) Then press the
ENT
key, and calibration
is performed in the selected range when
auto calibration or auto zero calibration is
performed.
Auto calibration and the manual calibration
of the component with which “AR” has
been selected as range switch mode are
performed in the range selected here. In this
case, once the calibration is started, the range
is automatically switched, and on completion
of the calibration, the original range is
resumed.
The range identification contact is inter-
locked with the range after the switch.
However, if the hold setting is set to “ON,”
the contact status before calibration is main-
tained.
“Auto Calibration Component/range”
setting
( )
ENT
(5) Press the key in the state described in
(3), and the highlight is switched between
“enable” and “disable” auto calibration.
(6) Select “enable” of “disable” by pressing
the or the key.
(7) Then press the
ENT
key.
( )
ENT
End of Auto Calibration
Range setting
End of Auto Calibration
component setting