6-11
Ver.1.36
6.2.5 List of recommended replacement parts
Pats listed below are consumable parts that are recommended to replace during maintenance work.
The replacement interval raries depending on the usage.
6.2.6 Periodic maintenance and inspection
We recommend that the pyrolyzer be overhauled and inspected every two years and that
hard-to-replace parts such as heater and thermal insulator be replaced for your stable and safety
operation of the product.
Device
Part name
Part No.
Replacement
cycle
Note
Pyrolyzer
(Furnace)
Quartz pyrolysis tube
A30
PY1-3018A
As needed
Replacement interval depends on type of
samples, amount used, and number of runs.
Can be heat-cleaned with torch flame.
Graphite Vespel ferrule
PY1-7911
As needed
Replace when pyrolysis tube M30 is replaced.
Needle set N
PY1-1274
As needed
Replacement recommended when pyrolysis
tube M30 is replaced.
Over tightening injector septum may clog
needle opening with septum debris.
ITF union N
PY1-3513
1 year
Standard accessory for EGA/PY-3030D
ITF union ASN
PY1-3533
1 year
Used when Auto-Shot sampler is used.
O-ring P-6W
PY1-2017
As needed
For sealing pyrolysis tube M30, 2 pcs used
Eco pickup
PY1-EP55F
1 year
Replacement not needed when Auto-Shot
Sampler is used all the time.
Septa for liquid
sampler, 20 pcs
PY1-2025
1 year
Inspection may only be needed if liquid sample
injection with micro syringe is not done.
O-ring P12
PY1-7811
1 year
For sampler base.
Inlet/outlet septum
PY1-2028
6 months
Used for carrier gas tube joints
Double-Shot
sampler
Cup chuck
PY1-1345
1 year
Worn-out chuck does not hold eco stick firmly.
Special tool required.
Overhaul
2 years
The product needs to be returned to F-Lab.
O-ring P-3
PY1-2112
6 months
GC
Absorption trap tube
PY1-2216
3~6 months
O-ring P-5
PY1-7814
3~6 months
Replace when trap tube is replaced. 4 pcs
used.
Septum
Agilent
PY1-7301
As needed
Replace if leakage detected when pyrolyzer
interface or GC injector cools down.
Replace when pyrolyzer is removed from GC
injector.
Shimadzu
PY1-7304
Others
Other parts that GC
manufacturers
recommend