Ver.1.36
5-11
5.6 SELECTION OF SAMPLE CUPS AND STICKS
The types of Eco-cups and Eco-sticks are selected from Table 5.1 depending on your analytical
purpose and the technique that you use.
Eco-cups and Eco-sticks are made of stainless steel; the entire surface has been deactivated by
coating with fused silica on the surface. The wall thickness is 0.1 mm for all.
Type / Appearance
Name
Part Number
Volume
(µL
)
Compatibility
with AS-1020E
Description
Eco-cups
Regular L
Eco-cup LF
PY1-EC80F
(100 pcs)
80
Yes
For all applications except for Micro UV
Irradiator
(UV/Py-GC/MS).
Optimum
pyrolysis temperature is 30ºC lower than
Eco-cup SF for most polymeric materials
due to large surface area. This is useful for
samples with relatively large volumes.
Regular S
Eco-cup SF
PY1-EC50F
(100 pcs)
50
No
For all applications except for Micro UV
Irradiator
(UV/Py-GC/MS).
This
cup
provides
slightly
better
pyrogram
reproducibility in flash pyrolysis-GC mode
than Eco-cup LF.
Flow through
Eco-cup
LHF
PY1-EC80HF
(20 pcs)
80
No
This cup provides better peak shape
resolution of volatile compounds (C3-10)
than regular cups. And it prevents
secondary reactions under a low carrier
gas flow, e,g. 10mL/min.
Side hole
Eco-cup UV
PY1-EC80UV
80
Yes
Used for analysis with Micro UV Irradiator
(UV/Py-GC/MS)
Eco-sticks
Short
Eco-Stick SF
PY1-ES10F
(50 pcs)
---
---
Used for manual Single-Shot, EGA or
Heart-Cut EGA analysis.
Long
Eco-Stick DF
PY1-ES20F
(50 pcs)
---
---
Used for manual Double-Shot analysis.
Table. 5.1 Selection of sample cups and sticks
4mm
8mm
5mm
4mm
1mm i.d.
hole
8mm
4mm
1mm i.d.
hole x 2
8mm
4mm
L=30 mm
L=80 mm