Operation of the pyrolyzer with the furnace temperature over 800°C for a long period time will
shorten the useful lifetime of the furnace heater. Operation of the pyrolyzer over 800°C should be 20
minutes or less. Also, when the pyrolyzer is at standby, keep the furnace temperature below 200°C.
This will extend the useful time of the furnace heater.
IMPORT ANT NOTE WHEN OPER ATING PYROLYZER OVER 800 °C
There is a risk of severe burn. Always ensure that the protective
housing cover is placed properly on the pyrolyzer. When removing
the cover from the pyrolyzer, make sure that the furnace
temperature is 100°C or below.
CAUTION
HOT
Risk of electrical shock. Do not remove the temperature controller
housing cover, unless you are a service engineer certified by
Frontier Laboratories Ltd..
WARNING
!
If dust is deposited on the power code plug or the cable
connectors, clean them thoroughly. Be sure to unplug the power
code. Do not use water or organic solvent. Use a dry cloth or
brush. There is a risk of fire, if dusty connectors are used.
WARNING
!
This product uses compressed air or nitrogen for cooling the
pyrolyzer furnace. Ensure that you have a good ventilation if
nitrogen gas is used. There is a risk of suffocation, if it is used in a
confined area with poor ventilation.
WARNING
!
When flammable solvents such as benzene, toluene, or acetone
are used for your sample, work in a well-ventilated area such as in
a hood and area where no open flame is used nearby to avoid
catching fire.
CAUTION
!
The interface needle of the pyrolyzer has a very sharp point. When
replacing the interface needle, be sure to wear protective gears
such as gloves, safety goggles, etc.
CAUTION
!
The pyrolyzer uses a quartz pyrolysis tube. If this tube is broken,
broken pieces have very sharp edges. When replacing the tube, be
sure to wear protective gears such as gloves, safety goggles, etc.
CAUTION
!