7-3
Ver.1.36
Quartz
pyrolysis
tube
is
broken
Tube cracks in lengthwise
direction.
The sampler base is fixed
with screws while the quartz
pyrolysis tube is not inserted
to the dead end of the
furnace.
Fixed the upper part in the state that a quartz
pyrolysis tube was not inserted in properly to
the depths and might tighten up ITF union. To
see an instruction manual, please confirm
whether a quartz pyrolysis tube is attached
definitely.
Tube cracks laterally.
Tube cracks at the welded
portion.
Misaligned blocks (interface,
furnace and sampler base).
Alignment blocks are required. Ask for repair.
7.2 TEMPERATURE CONTROLLER
SYMPTOM
CAUSE
SOLUTION/CHECK POINTS
Furnace will not heat
Heater setting is OFF.
Set it to ON (check the furnace box in the
configuration software)
Breaker is off.
Re-set the main breaker on the rear panel of
the controller.
Beep sound
Sensor cable is not connected
or contact failure
Ensure cable from pyrolyzer is properly
connected to connector labeled POWER,
SENSOR at rear of controller.
Sensor failure
Repair needed.
Cooling fan not running
It is blocked by a foreign
object
Make sure it is not blocked by foreign object or
gas tubing.
Fan failure
Repair needed.
Fan power not supplied
Ensure fan power cable is properly connected.
If it is, repair is needed.
Abnormal temperature
Sensor contact failure
Check to see if sensor cable from pyrolyzer is
properly connected.
Error in temperature
calibration
Redo calibration.
Sensor circuit failure
Repair needed.
Temperature unstable at low
temperatures (eg. 100ºC or
below)
ITF temperature may be too
high
Set ITF temperature control to Auto. (In Auto
mode, it is controlled to 100 ºC higher
temperature than pyrolyzer furnace.)