EtAL.9
Calibrating the upstream and downstream pressure
sensors
This menu records
the pressure value
in the EEPROM
according to the
selected tube.
The
Optima
can operate with various tube types. The choice of tube is made through:
!
the "MODE" key, set selection when the pump is on normal mode.
!
maintenance menu and SAV 2
The pressure calibration must be done for each type of tube active on the pump.
If a pressure sensor, the door sensor or CPU board has been changed, the calibration
values must be reset to zero for each tube. Where the CPU board has been changed, the
calibration must be carried out using maintenance software during the configuration
procedure.
Installation:
!
According to the diagram below:
The manometer must be placed at the same level as the Optima pump.
Pum p
M anom eter
T ap 3 w ays
C ontainer
Pock et
50 cm
1 m
Summary of Contents for OPTIMA PT-VS-S
Page 1: ...TECHNICAL MANUAL TECHNICAL MANUAL OPTIMA PT VS ST From Serial N 18180100 till ...
Page 6: ...66 Optima_PT VS ST_enTDM fm ...
Page 8: ...8 optima01_002b_en fm Presentation 1 2 Operation diagram ...
Page 20: ...20 optima02_002b_en fm Description and operation ...
Page 26: ...26 optima03 1_001b_en fm Description of the menus ...
Page 28: ...28 optima03 3_001b_en fm Description of the menus ...
Page 30: ...30 optima03 4_002b_en fm Description of the menus ...
Page 32: ...32 optima04 1 2_001b_en fm Preventive maintenance ...
Page 54: ...54 optima05 2_002a_en fm Troubleshooting ...
Page 56: ...56 optima06 0_001a_en fm Intervention procedures ...
Page 64: ...64 optima06 0_0020b_en fm Intervention procedures ...
Page 68: ...68 optima06 0_0030b_en fm Intervention procedures ...
Page 72: ...72 optima06 0_0040b_en fm Intervention procedures ...
Page 76: ...76 optima06 0_0050b_en fm Intervention procedures ...
Page 92: ...92 optima06 0_0110b_en fm Intervention procedures ...
Page 100: ...100 optima08 1_002b_en fm Spare parts catalogue ...
Page 104: ...104 optima08 3_001b_en fm Spare parts catalogue ...
Page 105: ......