Observations:
13
!
Check the upstream pressure. (Only
Optima VS
)
"
Start an infusion at 100 ml/h without drop detector.
"
Clamp the tube between the drop chamber and the pump.
"
Check that the
upstream
occlusion alarm is activated within 10 seconds.
"
Unclamp.
"
Check that the confirmation indicator lights up.
14
!
Check the clamp.
"
Place a water pipe in the device.
"
Close and open the door.
"
Check that the clamp lock the tube.
#
There should be no dripping at the tube outlet.
15
!
Check the battery operation.
"
Disconnect the mains lead.
"
Check the presence of the battery indicator.
"
Reconnect the mains lead.
"
Check if the battery indicator is switched off and that the mains indicator lights
up.
16
!
Check the flow rate.
Before any flow rate test, run an infusion of 20 ml to stabilise the flow rate.
"
Carry out a flow rate check in accordance with the norm EN 60601-2-24.
"
Check that the flow rate error is between - 5% and + 5%.
If there is too much discrepancy, refer to chapter 7 "Calibration / Calibration
procedure /
EtAL.2
calibrating the flow rate corrector coefficient".
17
!
Check the battery autonomy.
"
Start an infusion at 125 ml/h for 4 hours.
18
!
Carry out an electrical safety test according to norm EN 60601-1-2
N°
Procedure
Obtained value
Conformity
Yes
No
Name: Date:
Visa:
Summary of Contents for OPTIMA PT-VS-S
Page 1: ...TECHNICAL MANUAL TECHNICAL MANUAL OPTIMA PT VS ST From Serial N 18180100 till ...
Page 6: ...66 Optima_PT VS ST_enTDM fm ...
Page 8: ...8 optima01_002b_en fm Presentation 1 2 Operation diagram ...
Page 20: ...20 optima02_002b_en fm Description and operation ...
Page 26: ...26 optima03 1_001b_en fm Description of the menus ...
Page 28: ...28 optima03 3_001b_en fm Description of the menus ...
Page 30: ...30 optima03 4_002b_en fm Description of the menus ...
Page 32: ...32 optima04 1 2_001b_en fm Preventive maintenance ...
Page 54: ...54 optima05 2_002a_en fm Troubleshooting ...
Page 56: ...56 optima06 0_001a_en fm Intervention procedures ...
Page 64: ...64 optima06 0_0020b_en fm Intervention procedures ...
Page 68: ...68 optima06 0_0030b_en fm Intervention procedures ...
Page 72: ...72 optima06 0_0040b_en fm Intervention procedures ...
Page 76: ...76 optima06 0_0050b_en fm Intervention procedures ...
Page 92: ...92 optima06 0_0110b_en fm Intervention procedures ...
Page 100: ...100 optima08 1_002b_en fm Spare parts catalogue ...
Page 104: ...104 optima08 3_001b_en fm Spare parts catalogue ...
Page 105: ......