
5
Intended use
Volumat MC Agilia is an infusion pump designed for intravenous (IV) administration of drugs, solutions,
fluids, parenteral nutrition and transfusion (special set required). It must be used only by professionals (hospital
staff or caregivers).
Precautions to be taken
When a device is marked with the
symbol, operators must imperatively review the corresponding
Instructions for use prior to using this device. The use of infusion modes by untrained persons may lead to drug
administration errors.
Volumat MC Agilia has been tested in accordance with the electromagnetic compatibility standards
applicable to medical devices. Its immunity is designed to ensure correct operation. The limitation of the
emitted radiation avoids undesirable interference with other equipment such
as
EEG, ECG, etc. If Volumat MC
Agilia is placed near devices like HF surgical equipment, X-rays, NMR, mobile phones or Wi-Fi access points,
it is essential to observe a minimum distance between the Volumat MC Agilia and this equipment (see page 60
-Electromagnetic Immunity).
The device must not be used in
the
presence of inflammable anaesthetic agents due to a risk of explosion.
It should always be used away from all risk areas.
The device can be disturbed by pressure or pressure variations, mechanical shocks, heat ignition sources,
etc. If you wish to use the device in specific conditions, please contact our After-Sales Department. The pump
must be used in a horizontal and stable position to work correctly.
The physiological effects of medicine can be influenced by the characteristics of the device and the
associated disposable (constituent material is commonly listed on the set packaging). Check that they are
compatible with prescriptions, the characteristics of trumpet curves and occlusion alarm setting times in
relation to the programmed flow rate.
The device uses a Lithium-ion rechargeable battery. Incorrect handling of a Lithium-ion battery by non-
qualified personnel may cause battery leakage, overheating, smoke, explosion or fire, which could result in
deterioration of performance or failure. This may also damage the protection device installed in the battery pack,
resulting in damage to the equipment or injury to the user (see page 63 - Use of the internal battery).
In case of unexpected situation regarding pump controls or environment, the state of the art safe-design will
raise an alarm, stop the infusion and display an error code. Users are invited to be aware of those alarms (see
Chapter 6) and in cases where the device is used to deliver life sustaining therapies, like short half-life
medications, to consider adequate provisions for back-up therapy delivery solutions.
X
X
Programmed bolus
Bolus defined by its volume (or dose) and flow rate.
X
Loading dose
Initial dose defined by a time and delivered before a dose
rate.
X
Drops per minute
Infusion defined by a flow rate expressed in drops per
minute.
ml/h
Dose rate
Infusion mode
Description