Pressure history (in mmHg)
t
The date and user pressure limit is displayed on
the upper line.
u
The dotted curve represents the limit. The limit is
adjustable in User Menu [User 4: pressure] (see
page 46).
It is also adjustable during infusion in the Pressure
section of the menu.
v
The continuous curve represents the real pressure
during infusion.
The history is cancelled on drug change and
patient change.
The history runs over 2 hours.
Note
: On boluses and purge the pressure limit alarms are increased to their maximum level.
Example of an occlusion view
Example of the detailed screen of the event
This screen appears when pressing the eye key. User
limit indicates the limit set by the user. Current limit is
the pressure on infusion in the line.
Flow/Dose rate history
The history runs over 12 hours.
The upper line indicates the flow or dose rate.