47
Ward options are authorised options that you can select or not in User options (see previous table).
Option
Function
Choice
Ward
Ward code
Code:
(0200 by default)
Use increment and/or decrement keys, then
OK
for each digit
[Par 1] Beep sound
1 tonality
2 tonalities
Key beep
For preventive silence
Silence duration between 2 alarm beeps (0 to 5 seconds)
[Par 2] Sound level
7 sound levels available
[Par 3] Initial parameters
Drug and parameters
: last drug name and parameters are
displayed at start-up
Same infusion screen
: once activated, last infusion
parameters (VTBI, Time, Rate, VI) are recalled at switch ON for the
set duration
Volume infused
: deleted (VI=0 at switch ON),
stored (VI cumulated at switch ON)
[Par 4] Maximum rates
Choose for primary and secondary mode:
Macro infusion (ml/h)
Micro infusion (ml/h)
[Par 9] Bolus and
loading dose rates
Manual bolus
Maxi. macro (ml/h)
Maxi. micro (ml/h)
Storage: select option to store last bolus rates
Macro flow rate (ml/h)
Micro flow rate (ml/h)
Programmed bolus and
loading dose
Maxi. macro
Maxi. micro
[Par 10] Ward name
Press the increment and/or decrement keys to select
alphanumeric characters. Press
OK
after each selection.
[Par 11] Biomedical
name
[Par 12] User code
2-digit mandatory code to set code to user menu
[Par 13] Mains supply
disconnection alarm
Mains supply disconnection warning activated or not
[Par 17] Drug library
Selection of a drug library among four (maximum)
[Par 18] Night mode
Screen
brightness low
Green lights
low
Key beep off
Manual mode: manual
switch from one mode to
another
Auto mode: automatic switch
from one mode to another
according to the time range
settings
Select night mode. Use
the button to change mode.
From (hh:mm)
To (hh:mm)