6. Alarms and safety features
Volumat MC Agilia has a continuous inspection system that operates as soon as the pump is in use.
If an alarm occurs, the infusion stops, visual (red LEDs) and sound signals are emitted. A clear message is
expressed by means of words and pictograms. The
key is effective for two minutes. When the cause of the
alarm has been fixed, the red lights are turned off, but the message remains displayed at the top of the screen
as a reminder of the cause of the alarm.
If the pre-alarm or a warning occurs, the infusion continues, a visual (orange LEDs) and sound signals are
emitted. The
key is effective with no time limit.
System
Message
Type
Infusion
stop
Activation /
)
Action
Installed
set
Install set
Alarm
YES
At start-up: Infusion set not installed or door open
)
Install infusion set and close door.
Door opened
Alarm
YES
During infusion or stop mode: door is open.
)
Check set installation and close the door.
Set air installation
Alarm
YES
Tube is mis-positioned in front of air sensor.
)
Check set installation in front of air sensor and
close the door.
Air bubble
Alarm
YES
At start-up or stop mode: air bubble detected
)
Remove air bubble by priming set.
Air alarm
Alarm
YES
During infusion: air bubble detected.
)
Remove air bubble by priming set.
OCS
OCS failed
Alarm
YES
The OCS control system has detected a failure.
)
Check set installation, check door integrity,
check set integrity. If the problem cannot be solved,
contact the after-sales department.
Infusion
Flashing flow rate
Warning
NO
The flow rate has been modified from the keys but
has not been confirmed.
)
Check flow rate and confirm with
OK
.
End of infusion
pre-alarm
Pre-
alarm
NO
Remaining VTBI is less than 5% of initial VTBI set
up or 5 minutes or 5 ml left before initial VTBI is
reached.
)
Check if remaining volume in container is in
accordance with remaining VTBI. If needed,
prepare container for a new infusion sequence.
If operating with drop sensor, the end of the
infusion pre-alarm is inhibited and can be set in
ward option [Par 31].
End of infusion
alarm
Alarm
Stop/KVO
VTBI completed. KVO activated according to
configuration, see User option [User 5], page 46.
)
Press Stop to set new infusion parameters
(if required).