34
Changing an infusion set
1. Press the
key to hold the infusion.
2.
Close the roller clamp
.
3. Press the
key to silence the audible signal
for 2 minutes.
4. Open the pump door.
5. Disengage the infusion set from the pump.
6. Disconnect the infusion set from the container.
7. Disconnect the infusion set from the IV device in
accordance with local facility procedures.
8. Follow instructions as described from pages 11 to
13 (set installation and infusion adjustments).
Pre-programming the infusion
The Volumat MC Agilia can be pre-programmed
before installing the tubing set.
Switch on the device (door closed and without set)
and select the <
prog
> button.
Infusion adjustments can be made as described in
Operations on page 13.
When the parameters are logically entered and
confirmed by pressing
OK
, the
exit
and
C
options are
displayed as optional actions.
C
will allow parameter modifications and
exit
validates
the program parameters and the device will display
the "Install set" screen.
When the tubing set is installed, the device self -tests
and the programmed parameters are displayed.
Press
Start
to start the infusion or
C
to modify the
parameters.
Note
: The parameters are stored in the device and
are displayed when the machine is turned on.