48
Option
Function
Choice
Ward
[Par 19] Authorised
functions
Manual bolus
See air bubble
Prime set
Loading dose
Programmed bolus
[Par 20] Authorised units
Dilution units (Select/Deselect)
Dose rate units (Select/Deselect)
[Par 21] Mode displayed
Last mode: At switch ON, the
device will operate using the
previous programming mode
Question mode: At switch ON,
the device will ask which
programming mode is to be used
[Par 22] Authorised
modes
Drug labelling (or)
Vigilant
®
Drug’Lib
[Par 23] Patient
parameters
Select the patient's default parameters
[Par 24] Macro/Micro
mode
Macro: infusion
defined by increments
of 1 ml
Micro: infusion
defined by
increments of 0.1
ml
Storage: the last
choice is used at
switch ON
[Par 25] Same therapy
screen
'Same therapy' screen appears to restart infusion with last
parameters at the end of infusion. This function is available in Dose
Rate mode.
[Par 27] Air parameters
Macro parameters
(see below)
Micro parameters
(see below)
- Total volume/15 min (µl): above this volume of air, the alarm is triggered
(Adjustable from 10 to 2000 µl).
- Bubble filter (µl): Minimum bubble size taken into account (Adjustable from
0 to 250 µl).
[Par 28] Auto switch on at
door opening
Automatic switch on at door opening (select/deselect) when
device on mains supply
[Par 29] Infusion mode
Defines the available infusion modes in flow rate ml/h and in dose
rate
Volume/Time/Rate
Volume/Rate
Volume/Time
Time/Rate
Simple rate
Storage: stores the last
infusion mode
Drops/min
(in ml/h)
Ramp
(in ml/h)
Sequential
(in ml/h)
[Par 30] Drop sensor
Compulsory
(select/deselect)
For Primary/Secondary mode:
Drop sensor on primary (select/deselect)
Drop sensor on secondary (select/
deselect)