To set the measurement distance, please perform the following steps:
1. Click the measure distance set button.
2. According to the actual distance between the measured target and the
infrared thermal imager lens, slide the distance value on the screen up and
down, and set the distance value to the actual value.
3. Click on other areas of the thermal image screen, or click the back button to
complete the setting.
Note: The measurement distance is usually set to the default value at a short
distance or less than the longest accurate measurement distance of the thermal
imager.
4.6.6 Setting IR Window Compensation
IR window compensation is used to compensate the transmittance and
temperature of the external optics between measured object and the lens of the
thermal imager.
To set the IR window compensation, perform the following steps: