ROCK IMAGER User's Guide | Chapter 3: Specifications
Environmental Requirements
Ambient Temperature Range
0° C to 30° C.
Operating Humidity
0 - 80% non-condensing.
Storage Humidity
0 - 80% non-condensing.
Ventilation
At higher temperatures and humidity levels (e.g. >25°C and >65% humidity), condensation may
form on the outside of the Compressor-cooled RI 1000 when operated at or below 4° C. The
system will still maintain the desired temperature in these environments. However, the
compressor-cooled RI 1000 operates best in a temperature-controlled laboratory.
Store ROCK IMAGER within the specified ranges of temperature and humidity, in a
place where it is not exposed to direct sunlight.
ROCK IMAGER 1000
●
SQL Server Express/Standard Edition version 2008 R2 (Minimum) or 2014 (Recommended)
●
Dedicated database server (Recommended)
●
Network connection
●
Remote access (e.g. WebEx, TeamViewer, VPN) (Recommended)
●
Email Server access (SMTP server credentials)
●
Optional: UPS unit
●
Windows 7 32-bit or 64-bit, Windows 10 64-bit
●
4 GB RAM
●
2 GB available Hard Disk space
●
Dual core 1 GHz
●
1024 x 768 pixels screen resolution
●
3.5 and 4.5 .NET Framework
ROCK IMAGER 182/54
●
SQL Server Express/Standard Edition version 2008 R2 (Minimum) or 2014 (Recommended)
●
Dedicated database server (Recommended)
●
Network connection
Computer Requirements
Store ROCK IMAGER within the specified ranges of temperature and humidity, in a place
where it is not exposed to direct sunlight.
Do not block the fan vents on the ROCK IMAGER chassis.
Maintain at least 2 inches (50.8 mm) of free space between the
vents and any obstruction or surface.