FONA XPan 3D Operating Instructions  

10/28

 

 

69 683 70110 

-

 150831

 

3.3  Operating Positions 

 

 

PATIENT ENTRY 
position 

 

Control panel and X-ray 
source on the right of the 
patient and the image 

receiver on the left.  
 

 

 

 

END position 

 
At the end of the exposure 

the unit comes to a 
complete stop. 

 

 

START position 

 

System ready to start the 
exposure. When the unit 

reaches the START position 
the green light of the 

READY indicator on the 
control panel is turned ON 

 

 

PATIENT EXIT position 

 

Control panel and X-ray 
source on the left of the 

patient and the image 
receiver on the right.  

 

 

4. 

Accessories 

4.1  Rests and supports  

 

 

 

1.

 

Bite block with chin rest 

2.

 

Chin rest with support for edentulous  

 

 

 

 

3.

 

Bite block  

4.

 

Nasal support for edentulous patients  

Summary of Contents for XPan 3D

Page 1: ...FONA XPan 3D Operating Instructions English...

Page 2: ...ontained in this user manual Keep the machine in perfect working condition following the maintenance instructions given by the manufacturer Failure to observe the instructions relieves the manufacture...

Page 3: ...e show any changes in rated parameters or working ranges as well as the date the name of the firm and a signature For safety reasons only use original accessories indicated in this Operating Instructi...

Page 4: ...le of the release button permits in the designated significant zone of occupancy for the operator With the exception of the patient no other persons may remain in the room while the exposure is being...

Page 5: ...to Medical Devices and subsequent amendments and integrations of which in the Directive 2007 47 EC of the European Parliament and of the Council Nominal line voltage 230 V 10 115 V 10 Nominal line fre...

Page 6: ...eal size vertical magnification on adult standard profile 1 27 1 approximately Centering References Bite block Chin Rest for edentulous Aiming lights Type Class I LASER beam Wavelength 650 nm Output P...

Page 7: ...s OFF disconnected from mains supply Inherent Filtration ON connected to mains supply Fragile Handle With Care Fuse Fear of Humidity Alternate Current Up Do Not Overturn Protective Earth Stacking Limi...

Page 8: ...150831 3 Operating Controls and Displays 3 1 Unit 1 Main switch 2 Patient positioning mirror 3 Bite block 4 Image receptor 5 Height adjustment buttons 6 Knob for Frankfurt plane adjustment 7 Control p...

Page 9: ...tion Orange Light ON Key 4 PROGRAM Selection Key 2L 2R INCREASE kV left side mA right side Key 2L 2R DECREASE kV left side mA right side Key 13 PATIENT build Small Medium Large Extra Large Key 6 LIGHT...

Page 10: ...to a complete stop START position System ready to start the exposure When the unit reaches the START position the green light of the READY indicator on the control panel is turned ON PATIENT EXIT pos...

Page 11: ...he OrisWin DG Suite User Manual A Starting On the PC connected to FONA XPan 3D with OrisWin DG Suite installed Start OrisWin DG Suite and select the Patient module with the relevant button B Selecting...

Page 12: ...ical magnification on dental arch Program duration time approx 16 s Program exposure time 14 2 s The image at receptor s plane is approximately 27 higher than real size the vertical magnification on a...

Page 13: ...Program Left hemi arch Program duration time approx 14 s Program exposure time 7 3 s 6 4 P4 Program Right hemi arch Program duration time approx 16 s Program exposure time 7 3 s 6 5 P5 Program Anterio...

Page 14: ...nder the nose Once taken the first set of two images return the unit A second set of two exposures can be taken immediately TMJ closed mouth Program duration time approx 16 s Program exposure time 2 2...

Page 15: ...ons 69 683 70110 150831 15 28 6 8 P8 Program Full Mouth 3D Program exposure time 12 3 s 6 9 P9 Program available soon 3D TMJ left Program exposure time 12 3 s 6 10 P10 Program available soon 3D TMJ ri...

Page 16: ...d Start of the reset function has to be performed ATTENTION When switching on the unit there must NOT be a patient positioned in the unit If a fault occurs which requires switching the unit off and th...

Page 17: ...READY light ON TEST mode can be activated to run the unit without radiation On the display are shown The selected program number The message Test Mode instead of values of kV and mA A green light is a...

Page 18: ...motor starts slowly and then increases its speed Press the height adjustment key until the unit has reached the desired height Standard exposure program Bring the carriage to have the bite block or t...

Page 19: ...ad inclination for the FH setting by briefly touching the UP or DOWN height adjustment key Verify side rotation of the head with reference to the Midsagittal light using the mirror from the back of th...

Page 20: ...SE or DECRESE keys on control panel on unit or on PC side Upon manual correction of the pre set technique factors the corresponding light on the patient build is turned OFF Setting tube voltage is don...

Page 21: ...the patient ready to immediately interrupt radiation if necessary Expose the patient The exposure is released by keeping the exposure key pressed till end of movement The rotation movement runs automa...

Page 22: ...data acquisition program has to be started first to make sure data acquisition will be performed during radiographic exposure Automatic Exposure Blockage ALARM RED LIGHT BLINKING This feature protects...

Page 23: ...onditions By pressing the PROGRAM key select the exposure program for ADULT panorama P1 P3 P4 P5 P6 P7 or for CHILD panorama P2 or for 3D exam P8 P9 P10 By pressing the PATIENT key select the patient...

Page 24: ...ted DO NOT use solvents or corrosive substances The bite block and the chin rest can be sterilized in autoclave at 134 C 11 Inspection and maintenance Inspection and maintenance work must be performed...

Page 25: ...unit off Report to technical service 10 Tube head temperature exceeding limit Wait for tube cool down 11 Exposure aborted during irradiation Restart if termination was requested by the operator Call f...

Page 26: ...low The customer or the user of the FONA XPan 3D system should assure that it is used in such an environment Immunity test EN 60601 1 2 test level Compliance level Electromagnetic environment Electros...

Page 27: ...inity of equipment marked with the following symbol 13 4 Recommended safety distances for systems that are not life supporting The FONA XPan 3D system is intended for use in an electromagnetic environ...

Page 28: ...make any alterations which may be required due to clinical improvements FONA XPan 3D Operating Instructions English Edition 150831 FONA Dental s r o Stefanikova 7 SK 811 06 Bratislava Slovakia www fo...