FONA XPan 3D Operating Instructions 

69 683 70110

 

- 150831 

 

 23/28 

8. 

Programming 

Programming procedure 

In the factory kV and mA values are assigned to ADULT panoramic 
programs (P1, P3, P4, P5, P6, P7) and CHILD panoramic program (P2) 

for each one of the four types of patient build, for a total of 40 
combinations. 
Such values can be re-loaded using a dedicated service function 

 
Entering Service Mode 

 

By keeping pressed the BACK key for 5 s the system enters Service 
Mode 

 

Service function can be selected using  

the INCREASE or DECRESE keys on the RIGHT side of the display. 
Scroll the menu until Exposure Settings is displayed 

  Press the PROGRAM key to enter Exposure Settings. 

 

 NOTE. Keep pressed the TEST key to reset all values to default 

(factory) conditions  

 
 

By pressing the PROGRAM key select the exposure program  
for ADULT panorama (P1, P3, P4, P5, P6, P7) or  
for CHILD panorama (P2) or for 3D exam (P8, P9, P10). 

 

 

 

By pressing the PATIENT key select the patient build,  
Small, Medium, Large, Extra Large 

 

 

Change the kV level to the desired value by pressing the  
INCREASE or DECRESE keys at the LEFT side of the display 
Change the mA level to the desired value by pressing the  
INCREASE or DECRESE keys at the RIGHT side of the display 

 

Keep pressed the LIGHT

 

key to store the selected value 

  Repeat kV setting and mA setting  

for the other types of patient build if requested and for the other 
CHILD or ADULT group of programs if requested 

  Press BACK key to exit EXPOSURE SETTINGS 

 

Press TEST Mode key to exit SERVICE MODE 

 

 

Do verify proper setting by changing PATIENT build for ADULT 
panorama (programs P1, P3, P4, P5, P6, P7), CHILD panorama 

(program P2) and 3D (P8, P9, P10) 

 
 

* SET kV / mA * 
  67        6.3 

 S/- 
Expos. Settings

 

Summary of Contents for XPan 3D

Page 1: ...FONA XPan 3D Operating Instructions English...

Page 2: ...ontained in this user manual Keep the machine in perfect working condition following the maintenance instructions given by the manufacturer Failure to observe the instructions relieves the manufacture...

Page 3: ...e show any changes in rated parameters or working ranges as well as the date the name of the firm and a signature For safety reasons only use original accessories indicated in this Operating Instructi...

Page 4: ...le of the release button permits in the designated significant zone of occupancy for the operator With the exception of the patient no other persons may remain in the room while the exposure is being...

Page 5: ...to Medical Devices and subsequent amendments and integrations of which in the Directive 2007 47 EC of the European Parliament and of the Council Nominal line voltage 230 V 10 115 V 10 Nominal line fre...

Page 6: ...eal size vertical magnification on adult standard profile 1 27 1 approximately Centering References Bite block Chin Rest for edentulous Aiming lights Type Class I LASER beam Wavelength 650 nm Output P...

Page 7: ...s OFF disconnected from mains supply Inherent Filtration ON connected to mains supply Fragile Handle With Care Fuse Fear of Humidity Alternate Current Up Do Not Overturn Protective Earth Stacking Limi...

Page 8: ...150831 3 Operating Controls and Displays 3 1 Unit 1 Main switch 2 Patient positioning mirror 3 Bite block 4 Image receptor 5 Height adjustment buttons 6 Knob for Frankfurt plane adjustment 7 Control p...

Page 9: ...tion Orange Light ON Key 4 PROGRAM Selection Key 2L 2R INCREASE kV left side mA right side Key 2L 2R DECREASE kV left side mA right side Key 13 PATIENT build Small Medium Large Extra Large Key 6 LIGHT...

Page 10: ...to a complete stop START position System ready to start the exposure When the unit reaches the START position the green light of the READY indicator on the control panel is turned ON PATIENT EXIT pos...

Page 11: ...he OrisWin DG Suite User Manual A Starting On the PC connected to FONA XPan 3D with OrisWin DG Suite installed Start OrisWin DG Suite and select the Patient module with the relevant button B Selecting...

Page 12: ...ical magnification on dental arch Program duration time approx 16 s Program exposure time 14 2 s The image at receptor s plane is approximately 27 higher than real size the vertical magnification on a...

Page 13: ...Program Left hemi arch Program duration time approx 14 s Program exposure time 7 3 s 6 4 P4 Program Right hemi arch Program duration time approx 16 s Program exposure time 7 3 s 6 5 P5 Program Anterio...

Page 14: ...nder the nose Once taken the first set of two images return the unit A second set of two exposures can be taken immediately TMJ closed mouth Program duration time approx 16 s Program exposure time 2 2...

Page 15: ...ons 69 683 70110 150831 15 28 6 8 P8 Program Full Mouth 3D Program exposure time 12 3 s 6 9 P9 Program available soon 3D TMJ left Program exposure time 12 3 s 6 10 P10 Program available soon 3D TMJ ri...

Page 16: ...d Start of the reset function has to be performed ATTENTION When switching on the unit there must NOT be a patient positioned in the unit If a fault occurs which requires switching the unit off and th...

Page 17: ...READY light ON TEST mode can be activated to run the unit without radiation On the display are shown The selected program number The message Test Mode instead of values of kV and mA A green light is a...

Page 18: ...motor starts slowly and then increases its speed Press the height adjustment key until the unit has reached the desired height Standard exposure program Bring the carriage to have the bite block or t...

Page 19: ...ad inclination for the FH setting by briefly touching the UP or DOWN height adjustment key Verify side rotation of the head with reference to the Midsagittal light using the mirror from the back of th...

Page 20: ...SE or DECRESE keys on control panel on unit or on PC side Upon manual correction of the pre set technique factors the corresponding light on the patient build is turned OFF Setting tube voltage is don...

Page 21: ...the patient ready to immediately interrupt radiation if necessary Expose the patient The exposure is released by keeping the exposure key pressed till end of movement The rotation movement runs automa...

Page 22: ...data acquisition program has to be started first to make sure data acquisition will be performed during radiographic exposure Automatic Exposure Blockage ALARM RED LIGHT BLINKING This feature protects...

Page 23: ...onditions By pressing the PROGRAM key select the exposure program for ADULT panorama P1 P3 P4 P5 P6 P7 or for CHILD panorama P2 or for 3D exam P8 P9 P10 By pressing the PATIENT key select the patient...

Page 24: ...ted DO NOT use solvents or corrosive substances The bite block and the chin rest can be sterilized in autoclave at 134 C 11 Inspection and maintenance Inspection and maintenance work must be performed...

Page 25: ...unit off Report to technical service 10 Tube head temperature exceeding limit Wait for tube cool down 11 Exposure aborted during irradiation Restart if termination was requested by the operator Call f...

Page 26: ...low The customer or the user of the FONA XPan 3D system should assure that it is used in such an environment Immunity test EN 60601 1 2 test level Compliance level Electromagnetic environment Electros...

Page 27: ...inity of equipment marked with the following symbol 13 4 Recommended safety distances for systems that are not life supporting The FONA XPan 3D system is intended for use in an electromagnetic environ...

Page 28: ...make any alterations which may be required due to clinical improvements FONA XPan 3D Operating Instructions English Edition 150831 FONA Dental s r o Stefanikova 7 SK 811 06 Bratislava Slovakia www fo...

Reviews: