FONA XPan 3D Operating Instructions
26/28
69 683 70110
-
150831
13. Electromagnetic Compatibility
13.1 Electromagnetic emission
The FONA XPan 3D system is suitable for use in the electromagnetic environment specified below. The
customer or the user of the FONA XPan 3D system should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment
Radiated and conducted RF
emissions
CISPR 11
Group 1
The FONA XPan 3D system uses RF energy only for its
internal functions. Therefore, the RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
Class B
The FONA XPan 3D is suitable for use in domestic
environment and in environments directly connected to
the low voltage power supply network which supplies
buildings used for domestic purposes.
Harmonic Emissions
EN 61000-3-2
Not
applicable
Voltage fluctuations/ flicker
emissions 61000-3-3
Complies
13.2 Electromagnetic immunity
The FONA XPan 3D system is suitable for use in the electromagnetic environment specified below. The
customer or the user of the FONA XPan 3D system should assure that it is used in such an environment.
Immunity test
EN 60601-1-2
test level
Compliance level
Electromagnetic
environment
Electrostatic discharge (ESD)
EN 6 1000-4-2
6 kV contact
8 kV air
EN 60601-1-2
Test level
Residential /hospital
Radiated RF
EN 61000-4-3
Non-life-supporting equipment
3 V/m
80 Mhz to 2.5 GHz
Life-supporting equipment
10 V/m
80 MHz to 2.5 GHz
EN 60601-1-2
Test level
Residential /hospital
Conducted RF
EN 61000-4-6
Non-life-supporting equipment
3 V
150 kHz to 80 Mhz
Life-supporting equipment
3 V outside ISM band
10 V inside ISM band
EN 60601-1-2
Test level
Residential /hospital
Electrical fast transients/burst
EN 6 1000-4-4
2 kV for power supply lines
1 kV for input/output lines > 3 m
EN 60601-1-2 test
level
Residential /hospital
Surges
EN 61000 4-5
1 kV differential mode, 2 kV
common mode
EN 60601-1-2 test
level
Residential /hospital
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN 6 1000-4-11
0% U
n
for 0.5 cycle
40% U
n
for 5 cycles
70% U
n
for 25 cycles
0% U
n
for 5 sec
EN 60601-1-2 test
level
Residential /hospital
Power frequency (50/60 Hz)
magnetic field EN 61000-4-8
3 A/m
EN 60601-1-2 test
level
Residential /hospital